Description
∆T max. alarm
This alarm is triggered if the difference between Skin T1 and
Skin T2 is above the set value (risk of hypothermia).
Each of these alarms can either be set or switched in advance
to "OFF".
However, immediately after switching over to Kangaroo mode,
automatic alarm muting is activated for the first four minutes,
including for all special Kangaroo alarms, in order to give the
patient time to adapt to the new environment.
O
enrichment
2
When adjusting the oxygen concentration in the patient
capsule, the additional oxygen supply is metered by a micro-
processor-controlled valve.
The oxygen is thereby channelled into the air routing system,
so that it is heated and humidified with the air.