To remove the canopy:
1
Grasp the handles on the sides of the canopy with both
hands.
2
Lift the canopy horizontally off the pillar elements.
To replace the canopy:
2
Replace the canopy horizontally, so that the guide pins fit
into the holes in the pillar element.
Take care with the sensor unit!
Removing the bed
Maximum load must not exceed 5 kg
Open the front flap and fold it down.
l
Pull the bed out:
3
Turn both knobs to the vertical position marked
4
Pull the bed out towards the front as far as it will go, holding
it by the recessed handle or by the knobs.
When the bed is pulled out, the patient must be monitored
constantly to prevent the risk of a fall.
Do not lean or rest any weight on the bed when it is
pulled out.
After carrying out the care operations, push the bed back
l
until it clicks into place. Turn the knobs to the horizontal
position marked
, and close the front flap.
Push the bed all the way in! Otherwise the hot air duct will
be interrupted, and the patient may be warmed or cooled
excessively!