Preparation
D-Vapor with ISO conical adapter
without interlock system
1 Insert the vaporizer in the fresh-gas line.
2 On anesthesia workstations with rigid conical
connectors, adjustment plates (A) may be used
for alignment:
– Between the connecting piece and the
connecting plate and/or the connecting
plate and the anesthesia workstation
– Ensure adequate screw length – at least
4 thread pitches. If necessary, use longer
screws, strength at least 500 N/mm
3 Fasten the connecting plate (B) to the
connecting piece using M6 DIN 912-A4 screws;
tightening torque (7 ±0.5) Nm.
4 Tighten the vaporizer with a clamping plate and
2 M6 DIN 1587M-A4 cap nuts (C) and 2 A6.4
DIN 125-A4 washers.
WARNING
Risk of patient injury
Risk of incorrect output concentration and
high resistance may affect breathing.
The vaporizer is not suitable for use in a
breathing system.
50
A
B
A
C
2
.
D
5 Connect the gas inlet line and outlet line (D) to
the vaporizer.
WARNING
Risk of patient injury
The wrong flow direction will result in incor-
rect concentration.
When connecting the vaporizer, make sure
that the direction of flow is correct and corre-
sponds to the arrow on the back of the vapor-
izer, see page 39.
6 Press the 0 button (E) and set the control dial to
the 0 position.
CAUTION
Risk of injury and of damaging the vaporizer
A free-standing vaporizer must be secured
against tilting and falling.
WARNING
Mixtures, or concentrations that are too high
may be delivered.
Make sure that only one vaporizer can be used
at any one time or that only one vaporizer is
connected at any one time.
Instructions for use D-Vapor/D-Vapor 3000
E
0
D