Operation
Weighing scale
If the following instructions are not observed significant
weighing errors may occur:
Before the weighing process, make sure that both knobs are
l
set to the scales position
During the weighing process, Caleo
l
to any vibrations.
During the weighing process, no objects may be placed on
l
the bed surface.
During the weighing process, no objects may be placed
l
been the bed and the housing.
Hoses routed over the ventilation hose holder may affect the
l
accuracy of weighing.
The scales must be regularly tested with a test weight. In some
regions, the accuracy of the scales specified in the Technical
Data is only attained if the scales are calibrated at the installa-
tion location.
Preparing for weighing
®
Set Caleo
to the horizontal position. Check with the aid of
l
the built-in water levels – see page 27.
Remove all objects that touch both the bed and the fixed
l
environment or the incubator frame.
Set both knobs to the scales position
l
Prepare the patient for lifting.
l
Starting the weighing procedure
1
Activate the weighing process = Press button.
If the scale is not ready for operation, this function cannot
be activated.
88
.
®
must not be exposed
, see page 24.