Operation
Precautions
Weighing scale
Only use the built-in scales to determine the weight of the
patient.
If you fail to observe the Instructions for Use (see page 87),
considerable measuring errors may occur when determining
the weight. Consequently, to make sure that critical therapeutic
decisions based on patient weight are correct, the weight indi-
cated by the integrated scales must be checked against a
reference measurement carried out on external scales.
In-house transport
An even floor surface must be ensured when moving the
®
Caleo
incubator. The incubator must not be used outside the
hospital building, as the castors may be damaged or dislodged.
®
The Caleo
may only be moved when empty.
High noise levels
Excessive noise levels that can disturb the patient may be
caused by:
using O
head boxes and delivering pressurised gases,
2
wear on the bearings of the fan motor,
placing objects on the canopy.
Observe the specified maintenance intervals – see
l
page 106.
Cleaning mode
Cleaning mode should only be used when Caleo
®
After use, allow Caleo
to cool down before dismantling.
Risk of burning when touching the heater!
During or after cleaning mode, condensation could form under
the aggregate housing of the Caleo
42
®
is empty.
®
!
Electrical safety
Only use electromedical ancillary equipment conforming
to IEC / EN 60601-1.
Do not use multiway adapters to connect the power
®
supply of the Caleo
!
The patient leakage current may rise above the permitted
limits if the protective earth conductor fails while equipment is
connected to the socket strip.
In such cases, the risk of electric shock cannot be excluded.
When using the integrated socket strip:
take into account the total leakage current and total current
consumption!
See "Technical Data" on page 118.
Only connect external devices to the interfaces if Caleo
is connected to the power supply socket via the power
cable or the device is earthed via the earth connection on
the back of the unit.
Otherwise an electrical hazard could be possible.
Caleo does not monitor the power supply to external devices.