Dräger caleo Instructions For Use Manual page 99

Neonatal incubator
Hide thumbs Also See for caleo:
Table of Contents

Advertisement

1
Tip the bore sleeve to the side and remove it.
2
Tip the centering lugs to the side and remove them.
Remove visible soiling with a disposable cloth.
l
Wipe-disinfect the double wall surfaces.
l
After allowing the disinfectant time to take effect (see
l
manufacturer's specifications), wipe the surface with a clean
damp disposable cloth and then rub dry.
Clean all other parts in the cleaning and disinfecting
l
o
*
machine at 93
C
.
Removing the hood
3
Hold the hood with both hands using the handles on the
side.
4
Lift the hood vertically from the pillar elements.
Remove visible soiling with a disposable cloth and
l
detergent.
Wipe-disinfect the surfaces.
l
After allowing the disinfectant time to take effect (see
l
manufacturer's specifications), wipe the surface with a clean
damp disposable cloth and then rub dry.
Removing the hood support
The support (2M 51 141) is only mounted on the pillar
elements (2M 51 154 and 2M 51 156).
5
Move the support to a vertical position and remove it.
Clean the support in the cleaning and disinfecting machine
l
o
at 93
C.
*
Observe national and international standards on
cleaning, disinfecting and sterilisation procedures
(e.g. EN 285, EN 554, EN 556).
Disinfecting/Cleaning/Sterilising
1
2
3
4
Stripping down
3
4
5
99