Tilting the bed
To tilt the bed:
®
Switch on Caleo
(see page 43).
l
1
Press button = the bed will be lowered on the sensor
unit side.
2
Press button = the bed will be raised on the sensor unit side.
Adjust the bed to the required tilt angle.
l
The end position is reached when the tilt angle stops changing.
Release the button.
The entire housing of the Caleo
Hoses and cables must be carefully routed so that they do
not kink, tear or become squashed!
The height adjustment and tilt angle adjustment cannot be
operated at the same time!
Do not reach in between the housing and housing support
during adjustment. Danger of injury!
Setting the bed to the horizontal position
®
Caleo
must be switched on (see page 43).
l
1
Press button = the bed will be raised on the control unit side.
2
Press button = the bed will be lowered on the control unit
side.
The spirit levels show whether the bed is horizontal.
3
Spirit levels for the horizontal position of Caleo
transverse axis.
4
Spirit level for the horizontal position of Caleo
longitudinal axis. When using the integrated scales
(optional, see "Weighing scale" on page 87), make sure that
the unit is on a flat floor before setting.
®
incubator is tilted.
®
in the
®