Dräger caleo Instructions For Use Manual page 29

Neonatal incubator
Hide thumbs Also See for caleo:
Table of Contents

Advertisement

Integrated socket strip
The integrated socket strip can be used to connect
infusion pumps and
SpO
measuring equipment.
2
Connections may only be made by qualified technical staff.
Caleo does not monitor the power supply to external devices.
Do not exceed the maximum power input of the
connected accessories (all 4 sockets together:
max. 2 A).
The maximum permissible total leakage current must
not be exceeded.
For the leakage current of Caleo
strip, see "Technical Data" on page 118.
Connecting the nurse call
Socket on the rear of the control unit for connecting alarm
signals to a central alarm station in the hospital.
The kit may only be installed by specialists.
l
For details of the characteristics, refer to the Technical Data,
page 120.
The 6-pin round DIN plug (female connector) must be
l
connected to the lead for the central alarm station in the
hospital by a specialist.
Connection 3 to 5 will be closed and the nurse call is activated
in the event of an alarm with a high risk potential or if there is a
serious equipment fault.
The central hospital alarm system may only be
connected to the nurse call if Caleo is connected to the
mains power supply via a mains power cable or if it has
been earthed via the earth connection on the back of the
device.
Electric power may pose a hazard in all other cases.
1
Plug the connector into the »
control unit and screw into place.
Take care to route the connection cable so that damage
l
does not occur.
Ensure that the connector plug cannot be pulled out
l
accidentally.
Check correct operation of connected nurse call system.
l
*
optional equipment feature
®
without socket
*
« socket on the back of the
Preparation
Doors, ports and bed
1
5
3
1
29