Care
Assembling
Assembling
Mounting the expiration valve
The parts must be entirely dry to prevent malfunctioning.
Be careful to fit the diaphragm in the correct position.
●
Hold stopper by the flange and place diaphragm on the
collar of the stopper.
●
Insert stopper with diaphragm on top into the housing from
below and screw in tightly.
If the expiration valve has an additional water trap:
●
Fit collecting jar.
Medication nebuliser
●
Assemble in accordance with separate Instructions for Use.
●
Installation, see page 61.
Breathing gas humidifier
●
Assemble in accordance with separate Instructions for Use.
Before re-using on patient
●
Assemble machine as described under "Preparing for use"
on page 18 onwards.
●
Carry out checks to ensure readiness for operation,
see "Device Check" on page 30 onwards.