Technical Data; Quick Technical Data Guide; Applications; Function And System Design - Endress+Hauser Proline Prosonic Flow 93 Operating Instructions Manual

Foundation fieldbus ultrasonic flow measuring system
Hide thumbs Also See for Proline Prosonic Flow 93:
Table of Contents

Advertisement

Proline Prosonic Flow 93 FOUNDATION Fieldbus
Measuring principle
Measuring system
Measured variable
Measuring range
Operable flow range
Input signal
Endress+Hauser
10

Technical data

10.1

Quick technical data guide

10.1.1

Applications

• Measuring the flow rate of liquids in closed piping systems.
• Applications in measuring, control and regulation technology for monitoring processes.
10.1.2

Function and system design

The measuring system operates on the principle of transit time difference.
The measuring system consists of one transmitter and two sensors. A number of different versions
are available:
• Version for installation in the safe zone and for Ex Zone 2.
• Version for installation in Ex Zone 1 (see separate Ex documentation  ä 127)
Transmitter
Prosonic Flow 93
Sensor
• Prosonic Flow P Clamp-on version (for chemical and process applications),
nominal diameters DN 15 to 65 (½ to 2½")
• Prosonic Flow P Clamp-on version (for chemical and process applications),
nominal diameters DN 50 to 4000 (2 to 160")
• Prosonic Flow W Clamp-on version (water/wastewater applications),
nominal diameters DN 15 to 65 (½ to 2½")
• Prosonic Flow W Clamp-on version (water/wastewater applications),
nominal diameters DN 50 to 4000 (2 to 160")
• Prosonic Flow W Insertion version (water/wastewater applications)
nominal diameters DN 200 to 4000 (8 to 160")
• Prosonic Flow DDU 18 (sound velocity measurement),
nominal diameters DN 50 to 3000 (2 to 120")
• Prosonic Flow DDU 19 (wall thickness measurement),
– for wall thicknesses from 2 to 50 mm (0.08 to 2") for steel pipes
– for wall thicknesses from 4 to 15 mm (0.16 to ½") for plastic pipes
(only suitable for use with PTFE and PE pipes to a certain extent)
10.1.3

Input

Flow velocity
(transit time difference proportional to flow velocity)
Typically v = 0 to 15 m/s (0 to 50 ft/s)
Over 150 : 1
Status input (auxiliary input):
U = 3 to 30 V DC, R
= 5 k, galvanically isolated.
i
Configurable for:
Totalizer(s) reset, positive zero return, error-message reset, zero point adjustment.
Technical data
115