Installation
46
3.11.1
Installation for measurement as single-path insertion version
1.
Determine the installation area (e) on the pipe section:
– Mounting location ä 11
– Inlet/outlet run ä 12
– Space required by the measuring point: approx. 1× pipe diameter.
2.
Mark the middle line on the pipe at the mounting location and mark the position of the first
drillhole (drillhole diameter: 65 mm / 2.56").
!
Note!
Make the middle line longer than the drillhole!
e
Fig. 47:
Installing the measuring sensors, steps 1 and 2
3.
Cut the first hole, e.g. with a plasma cutter. Measure the wall thickness of the pipe if it is not
known.
4.
Determine the sensor distance.
!
Note!
Determine the sensor distance as follows:
– Via the Quick Setup "Sensor Installation" for measuring devices with local operation.
Run the Quick Setup as described on ä 84. The sensor distance is displayed in the
SENSOR DISTANCE function. The transmitter must be installed and connected to the
power supply before you can run the "Sensor Installation" Quick Setup.
– As explained on ä 84 for transmitters without local operation.
Fig. 48:
Installing the measuring sensors, steps 3 and 4
5.
Draw the sensor distance (a) starting from the middle line of the first drillhole.
Proline Prosonic Flow 93 FOUNDATION Fieldbus
A0001124