Operating Instructions
12
Technical data
12.1
Specifications
Raman Rxn2 analyzers may be configured to operate with one of several different laser wavelengths. Currently,
Raman Rxn2 analyzers may be equipped with a 532 nm, 785 nm, or 993 nm laser as standard.
12.1.1 Base unit
Item
Operating temperature
Storage temperature
Relative humidity
Warm-up time
Operating voltage
Transient over-voltages
Power consumption
Benchtop model dimensions (width x height x
depth)
Cart model dimensions (width x height x depth)
Weight
base unit
cart model
Sound level (from operator's perspective)
IEC 60529 rating
Altitude
Pollution degree
12.1.2 Spectrograph
Item
Type
Aperture ratio
Focal length
Grating (1 or 4 channels, 532 nm, 785 nm)
Grating (1 or 4 channels, 993 nm)
Spectral coverage (532 nm)
Spectral coverage (785 nm)
Spectral coverage: Raman Rxn2 hybrid
configuration (785 nm)
Spectral coverage (993 nm)
Slit
Spectral resolution (532 nm)
Spectral resolution (785 nm)
Spectral resolution (993 nm)
46
Description
15 to 30 C (59 to 86 F)
–15 to 50 C (5 to 122 F)
20 to 80 %, non-condensing
120 minutes
100 to 240 V, 50 to 60 Hz, ±10 %
Over-voltage category 2
400 W (maximum)
250 W (typical startup)
120 W (typical running)
279 x 483 x 592 mm
(10.99 x 19.02 x 23.31 in)
685 x 1022 to tabletop x 753 mm
26.97 x 40.24 to tabletop x 29.65 in
32 kg (71 lbs)
93 kg (205 lbs)
58.9 dB, A-weighted
IP20
Up to 2000 m
2
Description
Proprietary axial transmissive
f/1.8
85 mm
HoloPlex transmissive
(Raman Rxn2-785 hybrid analyzers use HoloSpec transmissive)
HoloSpec transmissive
150 to 4350 cm
-1
150 to 3425 cm
-1
175 to 1890 cm
-1
200 to 2400 cm
-1
50 μm fixed
5 cm
-1
average
4 cm
-1
average
-1
5 cm
average
Raman Rxn2
Endress+Hauser