List Of System Error Messages - Endress+Hauser Proline Prosonic Flow 93 Operating Instructions Manual

Foundation fieldbus ultrasonic flow measuring system
Hide thumbs Also See for Proline Prosonic Flow 93:
Table of Contents

Advertisement

Proline Prosonic Flow 93 FOUNDATION Fieldbus
No.
Error messages:
FOUNDATION Fieldbus
(FF)*
(local display)
* With FOUNDATION Fieldbus, error messages are displayed in the "Diagnosis" Transducer Block (base index: 1600) by means of the
"Diag. – Act.Sys.Condition" parameter (manufacturer-specific).
S = System error
$ = Fault message (with an effect on operation)
! = Notice message (without any effect on operation)
No. # 0xx  Hardware error
001
Device status message (FF):
ROM/RAM Failure –
Err. No. 001
Local display:
S: CRITICAL FAILURE
$: # 001
011
Device status message (FF):
Amplifier EEPROM failure –
Err. No. 011
Local display:
S: AMP HW EEPROM
$: # 011
012
Device status message (FF):
Amplifier EEPROM data
inconsistent –
Err. No. 012
Local display:
S: AMP SW EEPROM
$: # 012
041
Device status message (FF):
T-DAT failure –
Err. No. 041
Local display:
S: TRANSM. HW-DAT
$: # 041
042
Device status message (FF):
T-DAT data inconsistent –
Err. No. 042
Local display:
S: TRANSM. SW-DAT
$: # 042
Endress+Hauser
9.2.1

List of system error messages

Transducer Block
error messages
"Diagnosis" Transducer
Block
BLOCK_ERR = Device needs
maintenance now
Transducer_Error = Electronics
failure
BLOCK_ERR = Device needs
maintenance now
Transducer_Error = Data
integrity error
BLOCK_ERR = Device needs
maintenance now
Transducer_Error = Data
integrity error
BLOCK_ERR = Device needs
maintenance now
Transducer_Error = Electronics
failure
BLOCK_ERR = Device needs
maintenance now
Transducer_Error = Data
integrity error
Analog Input Function Block
error messages
OUT. QUALITY = BAD
OUT. SUBSTATUS = Device
Failure
BLOCK_ERR = Input Failure
(faulty input value from
Transducer Blocks)
OUT. QUALITY = BAD
OUT. SUBSTATUS = Device
Failure
BLOCK_ERR = Input Failure
(faulty input value from
Transducer Blocks)
OUT. QUALITY = BAD
OUT. SUBSTATUS = Device
Failure
BLOCK_ERR = Input Failure
(faulty input value from
Transducer Blocks)
OUT. QUALITY = BAD
OUT. SUBSTATUS = Device
Failure
BLOCK_ERR = Input Failure
(faulty input value from
Transducer Blocks)
OUT. QUALITY = BAD
OUT. SUBSTATUS = Device
Failure
BLOCK_ERR = Input Failure
(faulty input value from
Transducer Blocks)
Troubleshooting
Reason for error/rectification
(spare parts  ä 109)
Cause:
ROM/RAM error. Error when accessing the
program memory (ROM) or random access
memory (RAM) of the processor.
Remedy:
Replace the amplifier board.
Cause:
Amplifier with faulty EEPROM
Remedy:
Replace the amplifier board.
Cause:
Error when accessing data of the measuring
amplifier EEPROM
Remedy:
Perform a "warm start" (= start the measuring
system without disconnecting main power).
– FF: "Diagnosis" Transducer Block (base index:
1600)  "Sys. – Reset" parameter RESTART
SYSTEM
– Local display: SUPERVISION 
SYSTEM  OPERATION  SYSTEM RESET
( RESTART SYSTEM)
Cause:
1. T-DAT is not plugged into the amplifier board
correctly (or is missing).
2. T-DAT is defective.
Remedy:
1. Check whether the T-DAT is correctly plugged
into the amplifier board.
2. Replace the T-DAT if it is defective.
Check that the new, replacement DAT is
compatible with the measuring electronics.
Check the:
- Spare part set number
- Hardware revision code
3. Replace measuring electronic boards if
necessary.
4. Plug the T-DAT into the amplifier board.
101