Note!
The display returns to the cell SETUP SENSOR (1001) if you press the X key combination during parameter
interrogation. The stored parameters remain valid.
m
If a channel is selected for which a Quick Setup has already been executed, the previous values are overwritten.
n
During each run, all the options can be selected. If settings were made during a previous run, these are overwritten.
o
"Save?" prompt for pipe sound velocity::
• YES = The value measured during Quick Setup is accepted in the appropriate function.
• NO = The measurement is discarded and the original value remains.
p
The SOUND VELOCITY LINER (6529) only appears if:
• The LINER MATERIAL is selected to something ohter than NONE (6528).
q
The LINER THICKNESS (6530) only appears if:
• The LINER MATERIAL is selected to something ohter than NONE (6528).
r
The POSITION SENSOR function (6884) only appears if:
• The CLAMP ON option is selected in the MEASUREMENT function (6880)
and
• Two traverses are selected in the SENSOR CONFIGURATION function (6882)
s
The WIRE LENGTH function (6885) only appears if:
• The CLAMP ON option is selected in the MEASUREMENT function (6880)
and
• One traverse is selected in the SENSOR CONFIGURATION function (6882)
t
The ARC LENGTH function (6887) only appears if:
• The INSERTION option is selected in the MEASUREMENT function (6880)
and
• The DUAL-PATH option is selected in the SENSOR CONFIGURATION function (6882)
Commissioning
85