Endress+Hauser Proline Prosonic Flow P Brief Operating Instructions
Endress+Hauser Proline Prosonic Flow P Brief Operating Instructions

Endress+Hauser Proline Prosonic Flow P Brief Operating Instructions

Ultrasonic time-of-flight sensor
Hide thumbs Also See for Proline Prosonic Flow P:

Advertisement

Quick Links

KA01474D/06/EN/01.21
71508892
2021-05-10
Products
Brief Operating Instructions
Proline Prosonic Flow P
Ultrasonic time-of-flight sensor
These instructions are Brief Operating Instructions; they are
not a substitute for the Operating Instructions pertaining to
the device.
Brief Operating Instructions part 1 of 2: Sensor
Contain information about the sensor.
Brief Operating Instructions part 2 of 2: Transmitter
→  3.
Solutions
Services

Advertisement

Table of Contents
loading

Summary of Contents for Endress+Hauser Proline Prosonic Flow P

  • Page 1 Services KA01474D/06/EN/01.21 71508892 2021-05-10 Brief Operating Instructions Proline Prosonic Flow P Ultrasonic time-of-flight sensor These instructions are Brief Operating Instructions; they are not a substitute for the Operating Instructions pertaining to the device. Brief Operating Instructions part 1 of 2: Sensor Contain information about the sensor.
  • Page 2 Proline Prosonic Flow P Order code: XXXXX-XXXXXX Ser. no.: XXXXXXXXXXXX Ext. ord. cd.: XXX.XXXX.XX Serial number www.endress.com/deviceviewer Endress+Hauser Operations App A0023555 Endress+Hauser...
  • Page 3 These Brief Operating Instructions are the Brief Operating Instructionspart 1: Sensor. The "Brief Operating Instructions part 2: Transmitter" are available via: • Internet: www.endress.com/deviceviewer • Smart phone/tablet: Endress+Hauser Operations App Detailed information about the device can be found in the Operating Instructions and the other documentation: • Internet: www.endress.com/deviceviewer...
  • Page 4: Table Of Contents

    Table of contents Proline Prosonic Flow P Table of contents About this document ............. . 5 Symbols used .
  • Page 5: About This Document

    Proline Prosonic Flow P About this document About this document Symbols used 1.1.1 Safety symbols DANGER This symbol alerts you to a dangerous situation. Failure to avoid this situation will result in serious or fatal injury. WARNING This symbol alerts you to a dangerous situation. Failure to avoid this situation can result in serious or fatal injury.
  • Page 6: Basic Safety Instructions

    Basic safety instructions Proline Prosonic Flow P Symbol Meaning Protective Earth (PE) A terminal which must be connected to ground prior to establishing any other connections. The ground terminals are situated inside and outside the device: • Inner ground terminal: Connects the protectiv earth to the mains supply.
  • Page 7: Designated Use

    Proline Prosonic Flow P Basic safety instructions Designated use Application and media The measuring device described in these Operating Instructions is intended only for flow measurement of liquids. Depending on the version ordered, the measuring device can also measure potentially explosive, flammable, poisonous and oxidizing media.
  • Page 8: Operational Safety

    It meets general safety standards and legal requirements. It also complies with the EU directives listed in the device-specific EU Declaration of Conformity. Endress+Hauser confirms this by affixing the CE mark to the device. IT security Our warranty is valid only if the device is installed and used as described in the Operating Instructions.
  • Page 9: Product Identification

    • Enter serial numbers from nameplates in the W@M Device Viewer (www.endress.com/deviceviewer): All information about the device is displayed. • Enter the serial number from nameplates in the Endress+Hauser Operations App or scan the 2-D matrix code (QR code) on the nameplate using the Endress+Hauser Operations App: All information about the device is displayed.
  • Page 10: Storage And Transport

    Storage and transport Proline Prosonic Flow P Order code: Ser. no.: Ext. ord. cd.: Order code: Ser. no.: Ext. ord. cd.: A0030196  1 Example of a nameplate Order code Serial number (Ser. no.) Extended order code (Ext. ord. cd.)
  • Page 11 Proline Prosonic Flow P Installation 5.1.1 Mounting position Mounting location A0042039 Orientation Down Down A0041970  2 Orientation views Channel 1 upstream Channel 1 downstream Recommended orientation with upward direction of flow Non-recommended installation range with horizontal orientation (30°) Recommended installation range max. 120°...
  • Page 12 Installation Proline Prosonic Flow P Horizontal In the recommended installation range with a horizontal orientation (View B), gas and air accumulations at the top of the pipe and interference from deposit buildup at the bottom of the pipe can influence the measurement to a lesser degree.
  • Page 13 Proline Prosonic Flow P Installation 15 × DN 3 × DN 15 × DN 3 × DN 15 × DN 3 × DN 15 × DN 3 × DN 20 × DN 3 × DN 20 × DN 3 × DN A0042041 ...
  • Page 14 Installation Proline Prosonic Flow P 5.1.3 Special mounting instructions Weather protection cover: Proline 500 280 (11.0) 255 (10.0) 146 (5.75) 134 (5.3) 12 (0.47) 30 (1.18) A0029553  4 Weather protection cover for Proline 500; engineering unit mm (in) Cover locking NOTICE Order code for "Transmitter housing", option L "Cast, stainless": The covers of the...
  • Page 15: Mounting The Measuring Device

    Proline Prosonic Flow P Installation SW 2.5 5 (0.2) min. 15 (0.6) A0029799 Cover borehole for the securing screw Securing screw to lock the cover Mounting the measuring device 5.2.1 Required tools For transmitter Proline 500 For mounting on a post:...
  • Page 16 Installation Proline Prosonic Flow P 5.2.3 Mounting the measuring device ‣ Install the measuring device or turn the transmitter housing so that the cable entries do not point upwards.  A0029263 5.2.4 Mounting the sensor WARNING Risk of injury when mounting the sensors and strapping bands! ‣...
  • Page 17 Proline Prosonic Flow P Installation Screw the nuts onto the U-shaped screws. Position the sensor holder correctly and tighten the nuts uniformly. A0043369  5 Holder with U-shaped screws Sensor holder CAUTION Risk of damaging plastic or glass pipes if the nuts on the U-shaped screws are tightened too much! ‣...
  • Page 18 Installation Proline Prosonic Flow P Wrap the strapping bands around the sensor holder and measuring pipe without twisting them.  A0043371  6 Positioning the sensor holder and mounting the strapping bands Sensor holder Guide the strapping bands through the strapping band locks.
  • Page 19 Proline Prosonic Flow P Installation Push down the tensioning screw and tighten the strapping bands so they cannot slip.  A0043372  7 Tightening the tensioning screws of the strapping bands Sensor holder If necessary, shorten the strapping bands and trim the cut edges.
  • Page 20 Installation Proline Prosonic Flow P Set strapping band 1 to the desired position. Push down the tensioning screw and tighten strapping band 1 so it cannot slip. Strapping band 2: proceed as for strapping band 1 (steps 1 to 6).
  • Page 21 Proline Prosonic Flow P Installation Sensor holder with strapping bands (large nominal diameters) Can be used for • Measuring devices with measuring range DN 50 to 4000 (2 to 160") • Mounting on pipes DN > 600 (24") • 1-traverse mounting or 2-traverse mounting with 180° arrangement •...
  • Page 22 Installation Proline Prosonic Flow P 12. Only slightly tighten strapping band 2 for final mounting so that it can still be adjusted. The distance/offset from the center of strapping band 2 to the center of strapping band 1 is indicated by the sensor distance of the device.
  • Page 23 Proline Prosonic Flow P Installation *The distance between the strap bolt and strapping band lock must be at least 500 mm (20 in). • For 1-traverse mounting with 180° (opposite) (single-path measurement, A0044304), (two-path measurement, A0043168) • For 2-traverse mounting (single-path measurement, A0044305), (two-path measurement, A0043309) •...
  • Page 24 Installation Proline Prosonic Flow P Sensor installation – small nominal diameters DN 15 to 65 (½ to 2½") Requirements • The installation distance is known →  16 • The sensor holder is pre-assembled Material The following material is required for mounting: •...
  • Page 25 Proline Prosonic Flow P Installation Fit the sensor housing on the sensor holder.  A0043377  13 Fitting the sensor housing Endress+Hauser...
  • Page 26 Installation Proline Prosonic Flow P Lock the bracket in place to fix the sensor housing on the sensor holder.  A0043378  14 Fixing the sensor housing Connect the sensor cable to the adapter cable.  This completes the mounting procedure. The sensors can now be connected to the transmitter via the connecting cables.
  • Page 27 Proline Prosonic Flow P Installation Material The following material is required for mounting: • Two strapping bands incl. mounting bolts and centering plates where necessary (already pre-assembled →  19, →  21) • Two measuring wires, each with a cable lug and a fixer to fix the strapping bands •...
  • Page 28 Installation Proline Prosonic Flow P Take the still movable strapping band 2, incl. the mounting bolt, and move it until both measuring wires are evenly tensioned and then tighten strapping band 2 so that it cannot slip. Then check the sensor distance from the center of the strapping bands. If the distance is too small, release strapping band 2 again and position it better.
  • Page 29 Proline Prosonic Flow P Installation Attach the coupling pad with the adhesive side facing down on the sensors. Alternatively coat the contact surfaces with an even layer of coupling gel (approx. 1 mm (0.04 in)), going from the groove through the center to the opposite edge.
  • Page 30 Installation Proline Prosonic Flow P 10. Insert the sensor cable into the sensor until the end stop.  A0043383  19 Mounting the sensor and connecting the sensor cable The sensors can now be connected to the transmitter via the sensor cables and the error message can be checked in the sensor check function.
  • Page 31 Proline Prosonic Flow P Installation • Two sensors incl. connecting cables • Open-ended wrench (13 mm) • Screwdriver Procedure: Position the strapping bands using the mounting rail [only DN50 to 600 (2 to 24"), for larger nominal diameters, measure the distance between the center of the strap bolts...
  • Page 32 Installation Proline Prosonic Flow P Screw the mounting rail onto the sensor holders.  A0043385  21 Mounting the sensor holders and mounting rail Attach the coupling pad with the adhesive side facing down on the sensors (). Alternatively coat the contact surfaces with an even layer of coupling gel (approx.
  • Page 33 Proline Prosonic Flow P Installation 10. Insert the sensor cable into the sensor until the end stop.  A0043386  23 Mounting the sensor and connecting the sensor cable The sensors can now be connected to the transmitter via the sensor cables and the error message can be checked in the sensor check function.
  • Page 34 Installation Proline Prosonic Flow P Wall mounting 18 (0.71) 10 (0.39) 20 (0.79) 100 (3.94) A0029068  24 Engineering unit mm (in) Endress+Hauser...
  • Page 35 Proline Prosonic Flow P Installation Post mounting WARNING Order code for "Transmitter housing", option L "Cast, stainless": cast transmitters are very heavy. They are unstable if they are not mounted on a secure, fixed post. ‣ Only mount the transmitter on a secure, fixed post on a stable surface.
  • Page 36 Installation Proline Prosonic Flow P – – 3 mm 4 mm 7 Nm (5.2 lbf ft) A0029993  26 Non Ex housing Depending on the device version: Loosen the securing clamp of the connection compartment cover. Unscrew the connection compartment cover.
  • Page 37: Post-Installation Check

    Proline Prosonic Flow P Installation Turn the housing to the desired position. Tighten the securing screws. 5.2.7 Turning the display module The display module can be turned to optimize display readability and operability. – – – 3 mm A0030035 Depending on the device version: Loosen the securing clamp of the connection compartment cover.
  • Page 38: Disposal

    (WEEE), the product is marked with the depicted symbol in order to minimize the disposal of WEEE as unsorted municipal waste. Do not dispose of products bearing this marking as unsorted municipal waste. Instead, return them to Endress+Hauser for disposal under the applicable conditions.
  • Page 40 *71508892* 71508892 www.addresses.endress.com...