Proline Prosonic Flow 93 FOUNDATION Fieldbus
No.
Error messages:
FOUNDATION Fieldbus
(FF)*
(local display)
121
Device status message (FF):
Software compatibility
problem amplifier –
I/O module –
Err. No. 121
Local display:
S: A / C COMPATIB.
!: # 121
No. # 2xx  Error in DAT / no communication
205
Device status message (FF):
Save to T-DAT failed –
Err. No. 205
Local display:
S: LOAD T-DAT
!: # 205
206
Device status message (FF):
Restore from T-DAT failed –
Err. No. 206
Local display:
S: SAVE T-DAT
!: # 206
261
Device status message (FF):
Communication failure
amplifier –
Err. No. 261
Local display:
S: COMMUNICAT. I/O
$: # 261
Endress+Hauser
Transducer Block
error messages
"Diagnosis" Transducer
Block
BLOCK_ERR = Device needs
maintenance now
Transducer_Error = I/O failure
(input/output error)
BLOCK_ERR = Device needs
maintenance now
Transducer_Error = Electronics
failure
BLOCK_ERR = Device needs
maintenance now
Transducer_Error = Electronics
failure
BLOCK_ERR = Device needs
maintenance now
Transducer_Error = Electronics
failure
Analog Input Function Block
error messages
OUT. QUALITY = BAD
OUT. SUBSTATUS = Device
Failure
BLOCK_ERR = Input Failure
(faulty input value from
Transducer Blocks)
OUT. QUALITY = BAD
OUT. SUBSTATUS = Device
Failure
BLOCK_ERR = Input Failure
(faulty input value from
Transducer Blocks)
OUT. QUALITY = BAD
OUT. SUBSTATUS = Device
Failure
BLOCK_ERR = Input Failure
(faulty input value from
Transducer Blocks)
OUT. QUALITY = BAD
OUT. SUBSTATUS = Device
Failure
BLOCK_ERR = Input Failure
(faulty input value from
Transducer Blocks)
Troubleshooting
Reason for error/rectification
(spare parts  ä 109)
Cause:
Due to different software versions, I/O board and
amplifier board are only partially compatible
(possibly restricted functionality).
!
Note!
– This is indicated on the display as a warning
message for only 30 seconds (with entry in error
history).
– This situation in which the software versions
differ can occur if only one electronics board has
been exchanged; the extended software
functionality is not available. The previously
existing software functionality is still available
and measurement operation is possible.
Cause:
Data backup (download) to T-DAT failed, or error
when accessing (uploading) the calibration values
stored in the T-DAT.
Remedy:
1. Check whether the T-DAT is correctly plugged
into the amplifier board.
2. Replace T-DAT if faulty.
Before replacing the DAT, check that the new,
replacement DAT is compatible with the
measuring electronics. Check the:
– Spare part set number
– Hardware revision code
3. Replace measuring electronics boards if
necessary.
Cause:
Communication error. No data reception between
amplifier and I/O board or faulty internal data
transfer.
Remedy:
Check whether the electronics boards are correctly
inserted in their holders
103