3.6

Determining values for installation distances

3.6.1
Determining installation distances via local operation
Perform the following steps to determine the installation distances:
1.
Mount the wall-mount housing.
2.
Connect the power supply.
3.
Switch on the measuring device.
4.
Run the "Sensor Installation" Quick Setup menu.
Installing the wall-mount housing
There are various ways of installing the wall-mount housing:
• Direct wall mounting
• Panel mounting (with separate mounting kit, accessories  ä 92)
• Pipe mounting (with separate mounting kit, accessories  ä 92)
Caution!
• Make sure that the permitted operating temperature range
(-20 to +60 °C / –4 to +140 °F) is not exceeded at the mounting location. Install the device in a
shady location. Avoid direct sunlight.
• Always install the wall-mount housing in such a way that the cable entries are pointing down.
Direct wall mounting
Drill the holes  ä 17.
1.
2.
Remove the cover of the connection compartment (a).
3.
Push the two securing screws (b) through the appropriate bores (c) in the housing.
– Securing screws (M6): max. Ø 6.5 mm (0.26")
– Screw head: max. Ø 10.5 mm (0.41")
4.
Secure the transmitter housing to the wall as indicated.
5.
Screw the cover of the connection compartment (a) firmly onto the housing.
90 (3.54)
Fig. 10:
Direct wall mounting
35 (1.38)
b
c
a
Installation
c
192 (7.56)
mm (inch)
A0001130
17