Raman Rxn4 analyzers may be configured to operate with one of several different laser wavelengths. Currently,
Raman Rxn4 analyzers may be equipped with a 532 nm, 785 nm, or 993 nm laser as standard.
12.1.1 Base unit
Item
Operating temperature (532 nm, 785 nm)
Operating temperature (993 nm)
Storage temperature
Relative humidity
Warm-up time
Operating voltage
Transient over-voltages
Power consumption
Base unit dimensions (width x height x depth)
Weight, base unit
IEC 60529 rating
Altitude
Pollution degree
12.1.2 Base unit (w/NEMA 4x enclosure option)
Item
Operating temperature (532 nm, 785 nm, 993 nm)
Storage temperature
Relative humidity
Warm-up time
Operating voltage
Transient over voltages
Power consumption
Enclosure configuration with trolley option
(width  height x depth)
Weight, enclosure configuration with trolley option
IEC 60529 rating
50
Description
5 to 35 C (41 to 95 F)
5 to 30 C (41 to 86 F)
–15 to 50 C (5 to 122 F)
20 to 80 %, non-condensing
120 minutes
100 to 240 V, 50 Hz to 60 Hz, ±10 %
Over-voltage category 2
400 W (maximum)
250 W (typical startup)
120 W (typical running)
483 x 267 x 556 mm (19.02 x 10.52 x 21.89 in)
28.5 kg (63 lbs)
IP20
Up to 2000 m
2
Description
5 to 50 C (41 to 122 F)
–15 to 50 C (5 to 122 F)
Up to 80 % for temperature range 5 to 31 °C (41 to 87.8 °F), decreasing
linearly to 20 % at 50 °C (122 °F).
240 minutes
120 V ±10 %, 60 Hz or
230 V ±10 %, 50/60 Hz
Over-voltage category 2
1560 W (maximum)
1560 W (typical startup)
750 W (typical running)
1175 x 1480 x 826 mm (46.26 x 58.27 x 32.52 in)
185.5 kg (409 lbs)
IP66
Raman Rxn4