Technical Data - Endress+Hauser silopilot FMM 460 Installation And Operating Instructions Manual

Level measurement
Table of Contents

Advertisement

Silopilot FMM 4601460 Z
Technical Data
Final touches
Pull the cable glands tight and also seal them if necessary
(open at installation).
Carefully put the cover on the instrument and screw tightly.
If necessary, fit the protective sun cover and attach springs.
Set timers or adjust the digitallanalogue converter
+
memory. (see TI 11.73.14b)
Setting the timer (see TI 02.85.0111)
Set the timer to the required measuring frequency.
Refer to Fig. 15
Technical Data
Measuring range:
max. 30 m
Measuring error:
max. 1 pulse
Pulse output:
one pulse per 10 cm or per 1/10 foot
Other versions on request
Power supply:
refer to instrument versions/power supply
Heating voltage:
220V 50160 Hz or 1 10V 50160 Hz
Power consumption:
approx. 40 VA
Housing and flange:
Cast aluminium (G-AL)
Measuring tape FMM 460 Z:
Corrosion-resistant steel tape 0 1 mm
Option (not for Zone 10)
Oil-resistant nylon tape o 1 mm
Material of sensing weights:
refer to instrument versionslsensing weight
Max. gauge pressure p,:
2 bar in Silo
Protection conforming to DIN 40050:
Dust compartment FMM 460 IP 55
Dust compartment FMM 460 Z lP 65
Control compartment IP 55
Permissible temperature by instrument:
-1 5 O C
...
+
50° C (heating required below -10° C)
in vessel: -15O C ...+ 60° C
Mean tape speed:
approx. 18 cmls at 50 Hz (when raising or lowering)
Lifting force:
approx. 14 N, automatic motor shut-off with overstraining, automatic cut-in when
strain removed
Mounting:
flange to fit counter flange DN 100, PN 16 DIN 2633
Total weight:
approx. 15 kg