Installation
48
12. Check the distance between the drillholes once again and determine the path length.
!
Note!
Determine the path length as follows:
– Via the Quick Setup "Sensor Installation" for measuring devices with local operation.
Run the Quick Setup as described on  ä 84. The path length is displayed in the PATH
LENGTH function. The transmitter must be installed and connected to the power supply
before you can run the "Sensor Installation" Quick Setup.
– As explained on  ä 84 for transmitters without local operation.
13. Screw the sensors into the sensor holders by hand. If you use a tool, the maximum torque
permissible is 30 Nm.
14. Insert the sensor cable connectors into the opening provided and manually tighten the
connectors to the stop.
Fig. 52:
Installing the measuring sensors, steps 11 to 14
Proline Prosonic Flow 93 FOUNDATION Fieldbus
A0001129
Endress+Hauser