Troubleshooting
No.
Error messages:
FOUNDATION Fieldbus
(FF)*
(local display)
082
Device status message (FF):
Interruption between sensor
and transmitter CH1 -
Err. No. 082
Local display:
S: SENS. DOWN CH1
$: # 082
083
Device status message (FF):
Interruption between sensor
and transmitter CH2 -
Err. No. 083
Local display:
S: SENS. DOWN CH2
$: # 083
085
Device status message (FF):
Interruption between sensor
and transmitter CH1 -
Err. No. 085
Local display:
S: SENS. UP CH1
$: # 086
086
Device status message (FF):
Interruption between sensor
and transmitter CH2 -
Err. No. 086
Local display:
S: SENS. DOWN CH2
$: # 086
No. # 1xx  Software error
111
Device status message (FF):
Totalizer could not be restored
at startup –
Err. No. 111
Local display:
S: CHECKSUM TOTAL.
$: # 111
102
Transducer Block
error messages
"Diagnosis" Transducer
Block
BLOCK_ERR = Device needs
maintenance now
Transducer_Error =
Mechanical failure
BLOCK_ERR = Device needs
maintenance now
Transducer_Error =
Mechanical failure
BLOCK_ERR = Device needs
maintenance now
Transducer_Error =
Mechanical failure
BLOCK_ERR = Device needs
maintenance now
Transducer_Error =
Mechanical failure
BLOCK_ERR = Device needs
maintenance now
Transducer_Error = Electronics
failure
Proline Prosonic Flow 93 FOUNDATION Fieldbus
Analog Input Function Block
error messages
OUT. QUALITY = BAD
OUT. SUBSTATUS = Device
Failure
BLOCK_ERR = Input Failure
(faulty input value from
Transducer Blocks)
OUT. QUALITY = BAD
OUT. SUBSTATUS = Device
Failure
BLOCK_ERR = Input Failure
(faulty input value from
Transducer Blocks)
OUT. QUALITY = BAD
OUT. SUBSTATUS = Device
Failure
BLOCK_ERR = Input Failure
(faulty input value from
Transducer Blocks)
OUT. QUALITY = BAD
OUT. SUBSTATUS = Device
Failure
BLOCK_ERR = Input Failure
(faulty input value from
Transducer Blocks)
OUT. QUALITY = BAD
OUT. SUBSTATUS = Device
Failure
BLOCK_ERR = Input Failure
(faulty input value from
Transducer Blocks)
Reason for error/rectification
(spare parts  ä 109)
Cause:
Connection between sensor channel 1/2 and
transmitter interrupted.
Remedy:
1. Check the cable connection between the
sensor and the transmitter.
2. Check that the sensor connector is fully
screwed in.
3. Check that the correct sensor has been
connected.
4. Local display:
In the SENSOR TYPE function (6881), check
that the correct sensor has been selected for
Channel 1 or Channel 2.
5. FOUNDATION Fieldbus:
In the "Sensor Param. - Sensor Type"
parameter, check whether the correct sensor
has been selected for error number 82 or 83.
6. The sensor may be defective.
Cause:
Connection between sensor channel 1/2 and
transmitter interrupted.
Remedy:
1. Check the cable connection between the
sensor and the transmitter.
2. Check that the sensor connector is fully
screwed in.
3. Check that the correct sensor has been
connected.
4. Check that the correct sensor has been
selected.
– FF: Transducer Block "Transducer_CH1"
(base index: 1400) or Transducer Block
"Transducer_CH2" (base index: 1500)
Parameter "Sensor Param. - Sensor Type"
– Onsite display: BASIC FUNCTION 
SENSOR DATA CH1 or CH2 SENSOR
PARAMETERS SENSOR TYPE
5. The sensor may be defective.
Cause:
Checksum error of the totalizer.
Remedy:
1. Restart the measuring device.
2. Replace the amplifier board.