Endress+Hauser Proline Prosonic Flow 93 Operating Instructions Manual page 20

Foundation fieldbus ultrasonic flow measuring system
Hide thumbs Also See for Proline Prosonic Flow 93:
Table of Contents

Advertisement

Installation
!
20
Running the "Sensor Installation" Quick Setup menu
Note!
• If you are not familiar with the operation of the device  ä 69.
• The following section only describes the steps necessary for clamp-on and insertion type of
mounting within the "Sensor Installation" Quick Setup.
Running the Quick Setup for clamp-on type of mounting
1.
Enter or select installation-specific values or the values specified here.
2.
Read off the installation distances necessary for mounting.
Home position
Quick Setup
Clamp-on  Measurement
Subsequent procedure
The sensors can be installed once the installation distances have been determined:
• Prosonic Flow P (DN 15 to 65 / ½ to 2½")  ä 37
• Prosonic Flow P (DN 50 to 4000 / 2 to 160")  ä 37
• Prosonic Flow W  ä 41
Proline Prosonic Flow 93 FOUNDATION Fieldbus
 Setup sensor
Language
QS sensor units
CH1  Channel
Sensor type
1 or 2  Sensor configuration
Pipe standard
Nominal Diameter
Pipe material
Sound velocity pipe
Pipe diameter
Circumference
Wall thickness
Liquid
Temperature
Sound velocity liquid
Position sensor
Wire length
Sensor distance
No  Other measurement?
No  Other channel?
Setup sensor
= Number of traverses
Installation
Installation
distances for
distances for
measurement via
measurement via
one traverse:
two traverses: