Technical Data; Technical Data At A Glance - Endress+Hauser FEL 50 A Operating Instructions Manual

Table of Contents

Advertisement

Liquiphant M/S (HT), FEL 50 A PROFIBUS PA
Endress+Hauser
10

Technical data

10.1 Technical data at a glance

Application
Measuring principle
Measuring device
Measured variable
Fluid density
Output signal
Signal on alarm
Electrical connections
Cable entries
Supply voltage
Power consumption
Application
Maximum or minimum detection in tanks or piping with fluids of all types,
also in hazardous areas and in the food and pharmaceuticals industries.
Function and system design
The tuning fork of the sensor oscillates in resonance frequency. This
frequency is reduced when covered with liquid. The change in frequency
then activates a limit switch.
The device possesses a digital output with PROFIBUS PA protocol.
Input variables
Level (limit value)
Configuration using Commuwin II: > 0.5 g/cm
Output variables
• PROFIBUS PA
Setting
Level limit
GN
not inverted
inverted
Failure information can be opened using the following interfaces:
• Flashing yellow LED
• Status code
• Diagnosis code
Power supply
• 3 screw terminals
• PROFIBUS PA M12 plug-in connector
Cable gland: M20x1.5 or Pg 13.5
Cable entry: G ½ or ½ NPT
PROFIBUS PADiscrete-Input-Block M12 plug-in connector
Two-core cable connection, 9...32 V DC
100...350 mW
10 Technical data
3
3
or > 0.7 g/cm
FEL 50A
YE
OUT_D = 0
PA-Bussignal
OUT_D = 1
PA-Bussignal
OUT_D = 1
PA-Bussignal
OUT_D = 0
PA-Bussignal
37