Performance Specifications Agilent 1260 Infinity Standard Autosampler
(G1329B)
Specification
Operating range 0 - 60 MPa (0 - 600 bar, 0 - 8850 psi)
Early maintenance feedback (EMF), electronic records of maintenance and
errors
Controller-area network (CAN), RS232C, APG-remote standard, optional
four external contact closures and BCD vial number output
Leak detection and safe leak handling, low voltages in maintenance areas,
error detection and display
0.1 - 100 µL in 0.1 µL increments (recommended 1 µL increments)
Up to 1500 µL with multiple draw (hardware modification required)
1 – 99 from one vial
Typically < 0.25 % RSD of peak areas from < 5 µL to 100 µL
Typically < 1 % RSD of peak areas from 1 µL to 5 µL
1 µL from 5 µL sample in 100 µL microvial, or 1 µL from 10 µL sample in
300 µL microvial
Typically < 0.1 %, < 0.05 % with external needle cleaning
0.2 – 50 cp
100 × 2 mL vials in 1 tray
40 × 2 mL vials in ½ tray
15 × 6 mL vials in ½ tray (Agilent vials only)
50 s for draw speed 200 µL/min, ejection speed 200 µL/min, injection
volume 5 µL
Metering pump in high pressure flow path