Performance Specifications - Agilent Technologies 1200 Series User Manual

Thermostatted column compartment
Hide thumbs Also See for 1200 Series:
Table of Contents

Advertisement

2
Site Requirements and Specifications

Performance Specifications

Performance Specifications

Table 3
Type
Temperature
range
Temperature
stability
Temperature
accuracy
Column
capacity
Warm-up/co
ol-down time
Dead volume
Communicati
ons
Safety and
maintenance
28
Performance Specifications Thermostatted Column Compartment
Specification
10 degrees below ambient to 80 °C
10 degrees below ambient to 100 °C
up to 80 °C: flow rates up to 5 ml/min
up to 100 °C: flow rates up to
2.5 ml/min
± 0.15 °C
± 0.05 °C
± 0.8 °C
± 0.5 °C
Three 30 cm
5 minutes from ambient to 40 °C
10 minutes from 40 – 20 °C
3 µl left heat exchanger
6 µl right heat exchanger
Controller-area network (CAN), GPIB,
RS-232C, APG Remote: ready, start,
stop and shut-down signals, LAN via
other 1200 series module
Extensive diagnostics, error detection
and display (through control module
and Agilent ChemStation), leak
detection, safe leak handling, leak
output signal for shutdown of pumping
system. Low voltages in major
maintenance areas.
Comments
G1316A
G1316B (SL)/G1316C (SL Plus)
G1316A/G1316B (SL)/G1316C (SL Plus)
G1316B (SL)/G1316C (SL Plus)
G1316A
G1316B (SL)/G1316C (SL Plus)
With calibration
i.d. 0.17 mm, see
"Extended
Specifications on G1316B SL/G1316C SL
Plus"
on page 30
no GPIB on G1316B (SL)/G1316C (SL
Plus)
1200 Series TCC User Manual