Keithley 6220 Reference Manual page 388

Ac and dc current source measure unit
Table of Contents

Advertisement

17-12 Calibration
3.
4.
5.
6.
7.
8.
Figure 17-4
Connections for guard calibration
Connect
High and Low
of Output
Together
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176 - TestEquipmentDepot.com
Select GUARD, then press ENTER. The unit will display:
POSITIVE GUARD CAL
Press ENTER to Output +020.00uA
Press ENTER to output +20μA for positive guard calibration. The instru-
ment will prompt as follows:
DMM RDG=+22.00000mV
Use
,
,
,
, ENTER, or EXIT
Note the DMM reading, then adjust the Model 6220/6221 display to agree
with that value.
Press ENTER. The display will prompt:
NEGATIVE GUARD CAL
Press ENTER to Output -020.00uA
Press ENTER to output -20μA for negative guard calibration. The instru-
ment will prompt as follows:
DMM RDG=-22.00000mV
Use
,
,
,
, ENTER, or EXIT
Note the DMM reading, then adjust the Model 6220/6221 display to agree
with that value, and then press ENTER.
6220/6221
Output
Return to
Model 6220/6221 Reference Manual
Guard
Low
Section 17 topics
Digital Multimeter
(Measure DC Voltage)
High
Input
Low