Remote Wave Function Operation; Table 7-3 Waveform Example Parameters - Keithley 6220 Reference Manual

Ac and dc current source measure unit
Table of Contents

Advertisement

Model 6220/6221 Reference Manual

Remote wave function operation

Procedures for programming and generating waveforms for each of the four wave-
form types are given on the following pages. Each of these procedures includes
commands for a typical wave function example.
for each of these examples. See
page 7-27
Table 7-3
Waveform example parameters
Sine wave
Square wave
Ramp wave
Arbitrary wave
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176 - TestEquipmentDepot.com
"SCPI commands — wave functions," on
for details on wave function commands.
NOTE
User setups cannot be saved or recalled while
Wave is armed or running. The *SAV and *RCL
commands will generate error +413 Not al-
lowed with mode armed.
Waveform
(page
7-22)
(page
7-23)
(page
7-24)
(page
7-25)
Return to
Section 7 topics
Wave Functions (6221 Only) 7-21
Table 7-3
summarizes parameters
Parameters for waveform examples
Frequency: 1kHz
Amplitude: 10mA
Offset: 1mA
Phase marker: off
Duration: 5s
Frequency: 10kHz
Amplitude: 1mA
Offset: 0
Duty cycle: 50%
Phase marker: 180°, line 1
Duration: 10s
Frequency: 50kHz
Amplitude: 5mA
Offset: 0
Duty cycle: 25%
Phase marker: 90°, line 1
Duration: 15s
Arbitrary points: -0.5, -0.25, 0, 0.3, 0.4
Range: best fixed
Frequency: 100kHz
Amplitude: 25mA
Offset: 0
Phase marker: Off
Duration: 20s

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

6221

Table of Contents