4-14
Sweeps

Remote sweep operation

Procedures for programming and running a sweep for each of the three sweep types
are given on the following pages. Each of these procedures includes commands for a
typical sweep example.
ples. See
mands.
Table 4-3
Sweep example parameters
Linear staircase sweep
Logarithmic staircase sweep
Custom (list) sweep
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176 - TestEquipmentDepot.com
Table 4-3
"SCPI commands — sweeps," on page 4-18
NOTE
User setups cannot be saved or recalled while
sweep is armed or running. The *SAV and *RCL
commands will generate error +413 Not allowed
with mode armed.
Sweep type
(page
4-15)
(page
(page
4-17)
Return to
Section 4 topics
Model 6220/6221 Reference Manual
summarizes parameters for each of these exam-
for details on sweep com-
Parameters for sweep examples
Start current: 1mA
Stop current: 10mA
Step current: 1mA
Delay: 1s
Source range: best fixed
Compliance abort: off
Sweep count: 1
Bias current: 100μA
Compliance: 10V
4-16) Start current: 1mA
Stop current: 10mA
# points: 5
Delay: 2s
Source range: best fixed
Compliance abort: off
Sweep count: 1
Bias current: 10μA
Compliance: 5V
# points: 5
Points: 3mA, 1mA, 4mA, 5mA, 2mA
Delay: 3s, 1s, 4s, 5s, 2s
Compliance: 3V, 1V, 4V, 5V, 2V
Source range: auto
Compliance abort: off
Sweep count: 1
Bias current: 50μA