Keithley 6220 Reference Manual page 247

Ac and dc current source measure unit
Table of Contents

Advertisement

Model 6220/6221 Reference Manual
1. INIT
This command is used to initiate one trigger cycle and take the unit out of the
idle state.
2. ABOR
This command resets the trigger system. If a sweep is in progress, it will be
halted and the source value set to zero. The operate state, however, will be
left on. The sweep is still armed, and can be started again with another INIT
command (i.e., - no re-arming is needed).
3. ARM:SOUR <name>
This command selects the ARM layer event detector control source:
IMMediate, TIMer, BUS, TLINk, NSTest, PSTest, BSTest, or MANual.
4. ARM:SIGN
This command is used to bypass the ARM layer control source and move on
to the next layer in the trigger model (the TRIGger layer event detection in this
case).
5. ARM:TIM <n>
This command selects the ARM layer timer interval in the range of 0 to
99999.99 sec. with 1 msec resolution.
6. ARM:DIR <name>
This command enables (SOURce) or disables (ACCeptor) the ARM source
bypass.
7. ARM:ILIN <NRf>
This command selects the ARM layer input signal line: 1, 2, 3, 4, 5, or 6.
8. ARM:OLIN <NRf>
This command selects the ARM layer input signal line: 1, 2, 3, 4, 5, or 6. Note
that for the Model 6221 only, the trigger output line and phase marker
(SOUR:WAVE:PMAR:OLIN) lines cannot be the same, or a -221 Settings
Conflict error will occur. (See
9. ARM:OUTP <name>
This command selects when to output the ARM layer trigger. Available selec-
tions include TEXit (output when exiting trigger layer), TENTer (output when
entering trigger layer), or NONE (disabled).
10. TRIG:SOUR <name>
This command selects the TRIG layer event detector control source:
IMMediate and TLINk.
11. TRIG:SIGN
This command is used to bypass the TRIGger layer control source and move
on to the next layer in the trigger model (the device action block in this case).
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176 - TestEquipmentDepot.com
' Initiate one trigger cycle
' Reset trigger system
' Select ARM layer control source.
' Bypass ARM layer control source.
' Set ARM layer timer interval.
' Set ARM source bypass.
' Set ARM layer trigger input line.
' Set ARM layer trigger output line.
"Phase marker," on page
' Select when to output ARM trigger.
' Select TRIG layer control source.
' Bypass TRIG layer control source.
Return to
Section 8 topics
Triggering 8-11
7-9.).

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

6221

Table of Contents