8-10
Triggering
SCPI commands — triggering
Triggering commands are listed in
mand is provided in notes that follow the tables.
Table 8-2
Trigger commands
INITiate
ABORt
Arm layer commands:
ARM:SOURce <name>
ARM:SIGNal
ARM:TIMer <n>
ARM:DIRection <name>
ARM:ILINe <NRf>
ARM:OLINe <NRf>
ARM:OUTPut <name>
Trigger layer commands:
TRIGger:SOURce <name>
TRIGger:SIGNal
TRIGger:DIRection <name> Control TRIG source bypass.
TRIGger:ILINe <NRf>
TRIGger:OLINe <NRf>
TRIGger:OUTPut <name>
* Arm layer commands control entire sweep. Trigger layer commands control individual sweep steps.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176 - TestEquipmentDepot.com
Command*
Initiate one trigger cycle.
Reset trigger system.
Select event detector control source.
Bypass ARM control source.
Set timer interval (0 to 99999.99 sec).
Control ARM source bypass.
Set ARM layer input signal line (1-6).
Set ARM layer output signal line (1-6).
Enable/disable ARM layer output trigger.
Select event detector control source.
Bypass TRIG control layer.
Set TRIG layer input signal line (1-6).
Set TRIG layer output signal line (1-6).
Select TRIG layer output trigger.
Return to
Section 8 topics
Model 6220/6221 Reference Manual
Table
8-2. Additional information for each com-
Description
1
2
<name> = IMMediate, TIMer, BUS,
TLINk, NSTest, PSTest, NSTest,
or MANual.
5
1 msec resolution.
<name> = SOURce or ACCeptor
<name> = TENTer, TEXit, or NONE
<name> = IMMediate or TLINk.
<name> = SOURce or ACCeptor
<name> = SOURce[1], DELay, or NONE
Default
OFF
3
IMM
4
0.1
6
ACC
7
1
8
2
9
NONE
10
IMM
11
12
ACC
13
1
14
2
15
NONE