Model 6220/6221 Reference Manual
Calibration requirements
Be sure that you perform calibration:
•
•
•
•
•
Environmental conditions
Conduct the calibration procedures in a test environment that has:
•
•
Warm-up period
Allow the Model 622x to warm up for at least one hour before performing calibra-
tion.
If the instrument has been subjected to temperature extremes (those outside the
ranges stated above), allow additional time for the instrument's internal tempera-
ture to stabilize. Allow one extra hour to stabilize a unit that is 10°C (18°F) outside
the specified temperature range.
Also, allow the test equipment to warm up for the minimum time specified by the
manufacturer.
Line power
The Model 622x requires a line voltage of 100V to 240V and a line frequency of
50Hz to 60Hz.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176 - TestEquipmentDepot.com
Under the proper environmental conditions.
After the specified warm-up period.
Using the correct line voltage.
Using the proper test equipment.
Using the specified reading limits.
An ambient temperature of 18° to 28°C (65° to 82°F).
A relative humidity of less than 70% unless otherwise noted.
Return to
Section 17 topics
Calibration 17-3