Programming square waves

1.
2.
3.
4.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176 - TestEquipmentDepot.com
Restore defaults with this command:
*RST
Configure the waveform.
Examples – The following commands configure a 10kHz square wave with
an amplitude of 1mA, 0mA offset, 50% duty cycle, 180° phase marker, and
10 second duration:
SOUR:WAVE:FUNC SQU
SOUR:WAVE:FREQ 1e4
SOUR:WAVE:AMPL 1e-3
SOUR:WAVE:OFFS 0
SOUR:WAVE:DCYC 50
SOUR:WAVE:PMAR:STAT ON
SOUR:WAVE:PMAR 180
SOUR:WAVE:PMAR:OLIN 1
SOUR:WAVE:DUR:TIME 10
SOUR:WAVE:RANG BEST
Arm and trigger the waveform, turn on output:
SOUR:WAVE:ARM
SOUR:WAVE:INIT
To stop generating the waveform and turn the source output off before the
duration elapses, send this command:
SOUR:WAVE:ABOR
Return to
Wave Functions (6221 Only) 7-23
' Select square wave.
' Set frequency to 10kHz.
' Set amplitude to 1mA.
' Set offset to 0.
' Set duty cycle to 50%.
' Turn on phase marker.
' Set phase marker to 180°.
' Use line 1 for phase marker.
' 10 second duration.
' Select best fixed source range.
' Arm waveform.
' Turn on output, trigger waveform.
' Stop generating waveform.