Keithley 6220 Reference Manual page 117

Ac and dc current source measure unit
Table of Contents

Advertisement

Model 6220/6221 Reference Manual
Table 4-5
Custom (list) sweep commands
SOURce[1]:LIST:CURRent <NRf>
[,<NRf>, ...<NRf>]
SOURce[1]:LIST:CURRent:APPend <NRf>
[,<NRf>, ...<NRf>]
SOURce[1]:LIST:DELay <NRf>
[,<NRf>, ...<NRf>]
SOURce[1]:LIST:DELay:APPend <NRf>
[,<NRf>, ...<NRf>]
SOURce[1]:LIST:COMPliance <NRf>
[,<NRf>, ...<NRf>]
SOURce[1]:LIST:COMPliance:APPend <NRf>
[,<NRf>, ...<NRf>]
SOURce[1]:LIST:CURRent:POINts?
SOURce[1]:LIST:DELay:POINts?
SOURce[1]:LIST:COMPliance:POINts?
1. SOUR:LIST:CURR <NRf>[,<NRf>,...<NRf>]
SOUR:LIST:CURR:APP <NRf>[,<NRf>,...<NRf>] ' Append to current list
Use these commands to initially define a current list and to add current points
to an existing list. Note that the maximum number of sweep points is 64,000.
2. SOUR:LIST:DEL <NRf>[,<NRf>, ...<NRf>]
SOUR:LIST:DEL:APP <NRf>[,<NRf>, ...<NRf>] ' Append to delay list
These commands create a delay list and add delays to an existing list. Delay
values match up one-for-one with the current source values above in the cur-
rent list. These delays apply only to list (custom) sweeps. Note that defaults
are 1 second the same as regular source delay, and the minimum accepted
value is 0.001 seconds.
3. SOUR:LIST:COMP <NRf>[,<NRf>, ...<NRf>]
SOUR:LIST:COMP:APP <NRf>[,<NRf>, ...<NRf>] ' Append compliance list
These commands create a compliance list and add compliance values to an
existing list. Compliance values match up one-for-one with the current source
values above in the current list.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176 - TestEquipmentDepot.com
Command
Return to
Section 4 topics
Sweeps 4-21
Description
1
Defines list of currents.
<NRf> = -0.105 to 0.105 (A)
Adds current points to existing list.
<NRf> = -0.105 to 0.105 (A)
Defines list of delay values.
<NRf> = 0 to 999.9999 (s)
Adds to list of delay values.
<NRf> = 0.001 to 999999.999 (s)
Defines list of compliance values.
<NRf> = 0.1 to 105 (V)
Adds to list of compliance values.
<NRf> = 0.1 to 105 (V)
Query # of current list points.
Query # of delay list points.
Query # of delay list points.
' Define current list
' Define delay list
' Define compliance list
1
2
2
3
3
4
4
4