Model 6220/6221 Reference Manual

Restoring factory defaults

Before performing each procedure, restore the instrument to its factory defaults as
follows:
1.
2.

Test summary and considerations

Test summary

Model 622x test procedures include:

Test considerations

When performing the procedures:
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176 - TestEquipmentDepot.com
Press SETUP.
Select PRESET, then restore the factory default conditions by pressing
ENTER.
DC current output amplitude accuracy (Model 6220 and 6221),
Compliance accuracy (Model 6220 and 6221),
Waveform function amplitude and frequency accuracy (Model 6221 only),
page 16-9
Be sure to restore factory defaults as outlined above.
Make sure that the test equipment is properly warmed up and connected to
the correct Model 622x connector.
Be sure that a shorting connector is connected to the INTERLOCK connec-
tor, or you will not be able to turn on the output for testing. See Section 2 for
details on the interlock.
Make sure the test equipment and Model 622x settings are correct.
Return to
Performance Verification 16-5
page 16-8
Section 16 topics
page 16-6