8-16
Triggering
For this example, the Model 622x and switching mainframe are configured as follows:
Factory Defaults Restored
Trig-In Event = TLink
Trigger Input Line = #2
Trigger Output Line = #1
Trigger Output Delay Event = ON
Configure Sweep
To run this test properly, set up the sweep with 10 points with the desired currents
(see
Section 4
To start the test, press STEP on the switching mainframe to take it out of idle and
start the scan. The switching mainframes output pulse triggers the Model 622x to
output the sweep point current. The Model 622x then sends an output trigger
pulse to the switching mainframe to close the next channel. This process contin-
ues until all 10 sweep points are swept and channels are sequentially scanned.
Details of this testing process are explained in the following paragraphs and are
referenced to the operation model shown in
Figure 8-9
Operation model for triggering example
No
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176 - TestEquipmentDepot.com
Model 622x
(Section
4)
).
7001or 7002
Press STEP to start
Idle
Bypass
B
Wait for
Trigger Link
Trigger
Scan
C
Channel
Trigger
Output
D
Trigger
Scanned
10
Channels
?
Yes
Return to
Model 6220/6221 Reference Manual
Switching Mainframe
Factory Defaults Restored
Scan List = 1!1-1!10
Number of Scans = 1
Channel Spacing = TrigLink
Figure
8-9.
622x
Idle
A
Wait for
Trigger Link
Trigger
Output
E
Current
Trigger
Output
F
Trigger
Sourced
No
10
Points
?
Yes