Resource block error messages
Table 5-9. Resource Block BLOCK_ERR messages
Condition Name
Block configuration error
Simulate active
Power up
Out of Service
Device Fault State
5.5
Transducer block error messages
Error conditions that may appear in the Transducer block.
Table 5-10. Transducer Block BLOCK_ERR messages
Condition Name
Other error
Out of Service
96
Description
Configuration Error is used to indicate that you have
selected an item in FEATURES_SEL or CYCLE_SEL that
was not set in FEATURES or CYCLE_TYPE, respectively
This indicates that the simulation switch is in place. This
is not an indication that the I/O blocks are using
simulated data
The actual mode is Out Of Service
Set and cleared using SET_FSTATE and CLR_FSTATE
Description
Set whenever XD_ERROR is non-zero. See also
tools in AMS Device Manager" on page
The actual mode is out of service.