5.2.2

Viewing input and holding registers

Measured data is continuously stored in the Rosemount 2230's input registers. These registers can be
used for verifying that the Rosemount 2230 works properly and for advanced troubleshooting.
The Holding Registers store various configuration parameters which are used to control the display
presentation.
By using the TankMaster WinSetup configuration tool most holding registers can be edited by typing a
new value in the appropriate value input field.
To view input and holding registers:
1. Start the TankMaster WinSetup program
2. In the TankMaster WinSetup workspace window, click the right mouse button on the ATD device icon.
The ATD device represents all the non-level devices such as the Rosemount 2230 display.
3. Choose the View Input Registers option (or View Holding Registers), or from the Service menu
choose Devices > View Input / View Holding Registers.
Service and Troubleshooting
Service and Troubleshooting
July 2017
83