Recommended actions

The FD_RECOMMEN_ACT and RECOMMENDED_ACTION parameters display text strings that will give a recommended course of
action to take based on which type and which specific event of the alert that is active.
for Field Diagnostic Alerts as given by the alert default setting for the Rosemount 2230 Display.
Table 5-11. Recommended actions for Field Diagnostics Alerts.
Alert Type
Failure
Out of
Specification
Function
Check
Service and Troubleshooting
Description
Software Failure
Database Error
Electronics failure - Main Board
Electronics failure - FF I/O Board
Memory Failure - FF I/O Board
Internal Communication Failure
Internal Temperature Out of
Limits
Invalid Model Code
MAO Fault State Mode enabled
Check Function
Service and Troubleshooting
Table 5-11
Recommended Action
1. Restart communication.
2. Load default database to the device and reconfigure
the device.
3. Contact Rosemount Tank Gauging Service
Department.
1. Restart communication.
2. Load default database to the device.
3. Reconfigure the device.
Replace the device.
Replace the device.
1. Restart communication.
2. If error persists it may indicate a faulty memory chip.
Replace the device.
1. Restart communication.
2. Replace the device.
Check ambient temperature at installation site.
Contact Emerson Automation Solutions / Rosemount
Tank Gauging service department.
One or several Multiple Analog Output (MAO) blocks are
configured with fault-state mode enabled. With this
configuration, the output on the display will show the
fault-state data not the actual data from the connected
devices.
Turn fault state mode off in MAO block.
One or more transducer blocks are in Out of Service
mode. Return transducer block to Auto mode.