Replacing boards / replacing components
Pull the "Anode, E3 and Penning" cables leading from the high voltage section to the I.I.
out of the C-arm. (3/Fig. 2).
Loosen the retainer clip above the camera and remove it (2/Fig. 1 or 2/Fig. 4).
Remove the control panel of the I.I. mini voltage supply from the 9"/23cm I.I.
Loosen the camera set screw slightly and remove the camera (1/Fig. 3).
NOTE
Loosen the Compact optics (3/Fig. 3).
Remove the retaining clips on the cables (4/Fig. 1).
Remove the O-ring and the cap nut from the anode cable.
Disconnect plug M30.X1 (4/Fig. 3).
Remove the Compact optics.
To do this, pull the anode cable out of the Compact optics recess.
Rotate the C-arm so that the I.I. is underneath and the tube assembly above.
CAUTION
Remove the attachment screws from the I.I. mounting. When doing this, hold onto the I.I.
(5/Fig. 2).
Remove the I.I. from the mounting.

Installing the new I.I.

NOTE
Prior to installing the new I.I. on the C-arm mounting, complete the following work:
Reinstall the Compact optics on the new I.I.
- When doing this, feed the anode cable through the recess in the optics.
The Compact optics must sit adjacent to the eccentric screws.
Move the camera rotation of the Compact optics to the zero position.
- To do this, connect the ohmmeter to the potentiometer for camera rotation, position 2
and 3 (5/Fig. 3).
A resistance value of 5000 ohms, +40 ohms must be measured for the center position.
- Hold the camera rotation by the clamp ring (1/Fig. 3) and turn it to the correct position, if
necessary.
Siemens AG
Medical Engineering
Do not loosen the eccentric screws (Fig. 3/2) on the edge of the
Compact optics! These are used to center the Compact optics to
the I.I. output screen. When removing the Compact optics, make
sure that no dust or dirt contaminates the I.I. output window or
the Compact optics.
Risk of crushing!
Noncompliance may cause damage to property.
Apply all C-arm brakes and hold the I.I. when removing the attach-
ment screws.
When installing the Compact optics and the Videomed DC, make
sure that no dust or dirt contaminates the I.I. output window or
the Compact optics.
Register 5
RXR2-130.061.02
Rev. 03
Page 7 of 18
10.03
CS PS 24
6 - 7
6
SIREMOBIL Compact
System Manual