NOTE

analyzer location

Observe the following guidelines when selecting an analyzer installation
location:
1.
Select a readily accessible position for the analyzer to allow for routine
maintenance. Comfort levels for maintenance personnel should be
considered in placement of the sensor and control interface.
2.
The installation location should be free from excessive vibration and
the ambient temperature is required to be within the limits listed in
the specifications. If the ambient temperature is outside the speci-
fied limits or the vibration is excessive, please contact Thermox Sales
or Service Department at (412) 828-9040. We have special options to
address ambient temperatures outside the listed specifications and we
can provide special mounting solutions for excessive vibration installa-
tion locations.
InstallatIon
The operations in this chapter should be performed only by qualified
service personnel experienced in electrical safety techniques. There are
no operator-serviceable components inside the WDG system. Never
service the controller or sensor unless power has been removed from
the controller and sensor, and the sensor has been allowed to cool for
at least one hour. Also, always use gloves when working on the sensor.
Do not turn on the aspirator until the sensor has been turned on and
is hot - preferably 24 hours. If you turn off the analyzer or the process
is shut down, turn off the aspirator air to avoid plugging problems!
Installation | 3-1