System Messages; Auto Cal Pending; Auto Verify Pending; Calibration Aborted - Ametek Thermox Series 2000 WDG-IV User Manual

Table of Contents

Advertisement

System Messages

Message

Auto Cal Pending

Auto Verify Pending

Calibration Aborted

Calibration Required

Cal Verify Aborted

Power Down Detected

Primary Calibration

System Calibrating

System Verifying Cal

6-2 | Thermox Series 2000 WDG-IV / IVC / IVCM
Indicates that an automatic calibration is pending. once the current
calibration or verification is complete, the automatic calibration will
begin. This message will only appear if the Auto Calibration option
is installed and enabled.
Indicates that an automatic verification is pending. Once the current
calibration or verification is complete, the automatic verification will
begin. This message will only appear if the Auto Calibration option
is installed and enabled.
Indicates that the calibration has been aborted by the user, or the
calibration failed because the span gas or zero gas was out of
range.
Once you complete a primary calibration , this message is displayed
to inform you that a regular calibration is required (after a primary
calibration is performed, you must always then perform a regular
calibration). This message will not clear from the display by remov-
ing and restoring power to the control unit. If you see this message,
select the Calibrate key and perform a calibration.
The verification was successfully aborted by the system at your
request.
The system detected that power was removed, such as from a
power outage. Acknowledge this system message by setting or
reading any user-programmable value. This will clear the message
from the display. This message will not clear by removing and restor-
ing power to the unit.
Appears when the system is performing a Primary Calibration (see
the Setup key for further information on primary calibration). The
message continues to be displayed during the recovery period after
the primary calibration.
Indicates that a system calibration is in progress. It will appear
during the entire automatic calibration cycle or during the recovery
period of a remote or manual calibration.
Indicates that a system verification is in progress. It will appear
during the entire automatic verification cycle or during the recovery
period for a remote or manual verification.
Explanation