The following will then appear on the display:

Output Filtering Option

Filter out quick transitions in readings to allow a smoother analog output.
Setting output filtering
Select an analog output port.
From the Analog Range key, select Output Filtering. You will be
prompted to enter an output filtering value:
where XX is the current filter value.
100 equals no filtering, and 1 equals the highest possible filter value.
The default value is 100 (no filtering).
NOTE
Type a number between 1 and 100, then press Enter. Note that you
should enter 01 to enter 1, 07 to enter 7, 70 for 70, etc.
Out # 1 is set for
4 - 20 ma
o
r
Out # 1 is set for
0 - 20 ma
[100 =FAST
1 = SLOW
Fltr = XX New #?