To ensure a smooth installation of the AMETEK ta7000F Gas Purity Monitor and
to get the maximum performance, several aspects should be considered. This
guide discusses the important points for a successful installation.
PN T900-014, Rev B

Installation and Start-up

Before beginning the installation of the ta7000F and before powering
it up, review and follow all safety information following the Table of
Contents near the beginning of this manual. This information de-
scribes procedures to follow to avoid personal injury and/or damage
to the equipment. All regulatory agency and personnel safety proce-
dures for your jurisdiction must be followed.
The installation of the ta7000F must be in accordance with all of the
customer and local regulatory standards and procedures. There are no
operator-serviceable components inside the ta7000F. Refer all servic-
ing to qualified personnel.
Carrier Gas
The Flame Ionization Detector (FID) can be destroyed if operated at
temperature without flow. The carrier gas should remain flowing until
all heated zones are below 50 °C (2–3 hours).
Line Voltage
Do not connect the power cord to line voltage without checking
the correct operating voltage on the label located on the ta7000F Rear
Panel.
Actuator Gas Pressure
Always maintain correct actuator gas pressure. Low pressure (below
70 PSIG) can result in failure of the internal valves and/or stoppage
of the gas flow to sensitive instrument components. High pressure
(above 150 PSIG) will damage the valve actuators.
Installation and Start-Up
|
2-1