Overview; Aborting A Calibration - Ametek Thermox Series 2000 WDG-IV User Manual

Table of Contents

Advertisement

overview

aborting a Calibration

5-4 | Thermox Series 2000 WDG-IV / IVC / IVCM
CalibratE KEY
Menu options
Initiate Cal/Initiate Verify
Although automatic calibrations can be set to run at predetermined frequencies,
you can also force the system into an automatic calibration.
Cal/Verify Data
Cal Gas Values
Inject Cal Gas
Set Cal Timers/Set Auto Timers
(applies only for auto cal operations
with RCU, except for setting time and
date)
Set K Factor
(not used on WDG-IV Series)
Sensitivity
Pressing the Calibrate key while the system is performing a calibration
or verification, displays the option to abort the calibration or continue the
calibration. All other menu functions are temporarily disabled during a
calibration or verification. Oxygen, combustibles, and methane alarms are
also disabled during a calibration.
Functions/Options
Start an automatic, remote, or manual cali-
bration or verify
View latest calibration and verify data.
- Cal data
- Verify data
Define calibration/verify gas values to
match known values of calibration gas
cylinders.
Enter calibration gas values.
Set pressure and flow to RCU gases with-
out performing a calibration or verification.
Set control unit time/date.
Enable or disable auto cal cycles.
Set cal gas times for auto cal/verify opera-
tions.
Define recovery durations after a cal/verify.
Enter Value
View the Sensitivity (Slope) determined
during calibration for the combustibles
detector in mV/1000 PPM.