Select Function - Ametek Thermox Series 2000 WDG-IV User Manual

Table of Contents

Advertisement

Configuring the relay
1.
From the Alarm menu, select Relay Configure.
2.
Select De-energize on Alarm or Energize on Alarm.
If you select Energize on Alarm, the following message appears:
If you select De-energize on Alarm, the following message appears:

Select Function

Select which function Alarm 3 should track.
% Oxygen
When the system starts a calibration
When the system starts a verification
When the system starts a calibration or a verification (dual function)
When tracking the start of a calibration or verification, the alarm will de-
activate once the calibration or verification has been completed, including
any recovery time you have specified.
Defining the select function
1.
From the Alarm menu, select Select Function.
2.
Select Alarm3= Oxygen, Alarm3= In Cal., Alarm3= In Verify, or
Alarm3= Cal/Verify, depending on whether Alarm 3 should be associ-
ated with % oxygen levels, or with the start of a calibration or verifi-
cation, or with the start of a calibration or verification, or both (dual
function).
The control unit automatically exits from this menu once you make your
selection.
Relays are set to
ENERGIZE on Alarm
Relays are set to
DE-ENERGIZE on Alarm
Controller / User Interface | 4-25

Advertisement

Table of Contents
loading

Table of Contents