Always use tubing that is free of oil and dirt. Stainless steel tubing is rec-
ommended.
Figure 3-8. Z-purge setup.
Z-Purge Connections
Connect instrument air to the sensor Z-purge inlet. Maintain the pressure
and flow as directed on the Z-purge warning label.
You must provide a protective gas supply with an alarm for a loss of
pressure indication in order to meet NFPA 496 requirements.
You must not exceed the maximum temperature rating listed in the
specifications.
Process pressure is assumed to always be subatmospheric (vacuum)
unless the High-pressure option is installed.
The WDG-IV sensor contains a flame arrestor to prevent it from being
an ignition source to the process for short periods of high combustible
levels (25% of LEL - lower explosive limit) in the process. The flame
arrestor is not, however, intended to protect the process where the
combustibles levels are always high.
Installation | 3-13