Sensor : Wdg-Ivcm - Ametek Thermox Series 2000 WDG-IV User Manual

Table of Contents

Advertisement

Sensor : WDG-IVCM

Output Range
Oxygen
Combustibles
Methane
Accuracy
Oxygen
Combustibles
Methane
Response O2
Drift
Aspirator Air
Requirements
Max. Flue Gas Temp / Probe Type / Lengths
1300°F (704°c)
1875°F (1024°c)
3000°F (1649°C)
Max. Sample
Dewpoint
Sample Pressure
Environment
Ambient Temp.:
Relative Humidity: 10% to 90%, non-condensing
Enclosure
Lift-off NEMA 3R, weather resistant, stainless steel. Optional hinged NEMA 4X (IP56), explo-
sion-proof, purged, and floor mount versions available.
Power Requirements
2-4 | Thermox Series 2000 WDG-IV / IVC / IVCM
From 0-1% to 0-100%
From 0-1% to 0-100%
From 0-2,000 ppm to 0-10,000 ppm
0 to 5%
± 0.75% of measured value or ± 0.05% oxygen, whichever is greater
± 2% of full scale output range
±5% of full scale output range
90% of a process step change < 13 secs.
< 0.1% of cell output per month (< 0.005% O
applied)
10 to 20 scfh (4.72 to 9.4 L/min) at 15 to 100 psig (1.05 to 7.04 kg/cm
316 ss
36" - 108" (91 cm - 271 cm)
310 ss
36" - 108" (91 cm - 271 cm)
®
Hexoloy
24" - 72" (60 cm-182 cm)
392°F (200°C)
± 10 inch water gauge
-5°F to 160°F (-20°C to 71°C);
-5°F to 140°F (-20°C to 60°C) with Div. 2 Option
115 VAC, ± 10%, 47-63 Hz, 700VA max.
(750 VA max. floor mount option)
per month with 2% O
2
230 VAC, ± 10%, 47-63 Hz, 1950 VA max.
(2000 VA max. floor mount option)
2
2
)

Advertisement

Table of Contents
loading

Table of Contents