Overview Of Alarm Functions - Ametek Thermox Series 2000 WDG-IV User Manual

Table of Contents

Advertisement

Overview of Alarm Functions

Define whether Alarm 3 activates based on oxygen levels, the start of a calibration, or
the start of a verification (Select Function). All other alarms are set to activate based on
only a single condition ; for Alarm 4 this condition is levels of oxygen, for Alarms 5 and
6 this condition is high combustible levels (% or PPM scale), for Alarms 7 and 8 this
condition is high methane.
*Select Function
Alarm 3
% O
2
Cal Start
Verify Start
Cal/Verify Start
Alarm 4
Not selectable
(% O
only)
2
Alarm 5
Not selectable
Alarm 6
(combustibles only)
Alarm 7
Not selectable
(methane only)
Alarm 8
*
The Select Function menu option applies only to Alarm 3. All other alarms are restricted
to activate based on process reading levels.
Alarm Set Points
Exception Log
Alarms 1 and 2 are reserved for Watchdog and Service alarms and are not software-
selectable. They always de-energize on alarm (FAIL SAFE).
Alarm Configure
Relay Configure
High O
Alarm
2
Energize on alarm
Low O
Alarm
2
De-energize on alarm
Not Selectable
(high alarms only)
(Disabled during
calibration cycle)
For oxygen alarms, these values depend on the Alarm
Configure menu option for whether they are high or low
alarms.
The exception log provides a history of events, such as
system or error messages, alarm conditions, and the
start of calibrations or verifications.
Controller / User Interface | 4-21