Table of Contents

Advertisement

Quick Links

Thermox
WDG-V
®
RM
User Manual
Thermox
150 Freeport Road
Pittsburgh, PA 15238
9000-166-VE REV H

Advertisement

Table of Contents
loading

Summary of Contents for Ametek Thermox WDG-VRM

  • Page 1 Thermox WDG-V ® User Manual Thermox 150 Freeport Road Pittsburgh, PA 15238 9000-166-VE REV H...
  • Page 2: Offices

    If the instrument or procedures are used for purposes over and above the capabilities specified herein, confirmation of their validity and suitability should be obtained; otherwise, AMETEK does not guarantee results and assumes no obligation or liability. This publication is not a license to operate under, or a recommendation to infringe upon, any process patents.
  • Page 3: Table Of Contents

    Contents OFFICES ....................... ii Safety Notes ......................v Important Notice to Users ..................vi Environmental Information (WEEE) ..............vii General Safety Summary ..................viii Product Damage Precautions ................viii Declaration of Conformity ..................ix Warranty and Claims ....................x CHAPTER 1 SYSTEM OVERVIEW Sensor Operations ....................1-1 Common Operator Errors ...................
  • Page 4 Sensor Configuration ..................5-1 Analog Outputs ....................5-2 Setting the Alarms ....................5-5 Communications ....................5-7 System Time ....................... 5-9 Passcode Configuration ..................5-10 System Name Configuration ................5-12 Flow Meter Configuration ................5-13 Flow Calibration Procedure ................5-14 CHAPTER 6 CALIBRATION Types of Calibrations ..................
  • Page 5: Safety Notes

    Safety Notes WARNINGS, CAUTIONS, and NOTES contained in this manual emphasize critical instructions as follows: An operating procedure which, if not strictly observed, may result in personal injury or envi- ronmental contamination. An operating procedure which, if not strictly observed, may result in damage to the equipment. Important information that should not be overlooked.
  • Page 6: Important Notice To Users

    AMETEK recom- mends that all equipment requiring service be sent back to the factory. You should only attempt to repair or service this equipment after receiving training from an AMETEK/P&AI Division training representative.
  • Page 7: Environmental Information (Weee)

    Environmental Information (WEEE) This AMETEK product contains materials that can be reclaimed and recycled. In some cases the product may contain materials known to be hazardous to the environment or human health. In order to prevent the release of harmful substances into the environment and to conserve our natural resources, AMETEK recom- mends that you arrange to recycle this product when it reached its “end of life”.
  • Page 8: General Safety Summary

    General Safety Summary Review the following safety precautions to avoid injury and prevent damage to this product or any products connected to it. Use Proper Wiring To avoid fire hazards, use only the wiring specified in the Installation Chapter of this user’s manual. Avoid Electrical Overload To avoid electrical shock or fire hazard, do not apply a voltage to a terminal that is outside the range specified for that terminal.
  • Page 9: Declaration Of Conformity

    CAN/CSA 22.2 No. 1010.1-92 cUL Listed Inspection and Measuring Electrical Equipment Low Voltage Directive 73/23/EEC: EN61010-1:1993, Amendment A2: 1995 Electrical Equipment for Measurement, Controls, and Laboratory Use Ingress Protection: IP65 Manufacturer’s Address in Europe: AMETEK Precision Instruments Europe GmbH Rudolf-Diesel-Strasse 16 D-40670 Meerbusch, Germany June 2013 Contents | ix...
  • Page 10: Warranty And Claims

    WARRANTY AND CLAIMS We warrant that any equipment of our own manufacture or manufactured for us pursuant to our specifications which shall not be, at the time of shipment thereof by or for us, free from defects in material or workmanship under normal use and service will be repaired or replaced (at our option) by us free of charge, provided that written notice of such defect is received by us within twelve (12) months from date of shipment of portable analyzers or within eighteen (18) months from date of shipment or twelve (12) months from date of installation...
  • Page 11: Chapter 1 System Overview

    SYSTEM OVERVIEW Sensor Operations Designed for fast response in a Continuous Emissions Monitoring System, the sensor provides continuous measurement of oxygen. A sample is directed from the process stream to the analyzer and then returned to the process. A portion of this gas rises past the oxygen measuring cell and returns to the primary loop.
  • Page 12 Sensor Basic Elements of the Sensor The WDG-V series analyzer consists of the following basic systems: The Plumbing  All inlet and outlet tubing (cell housing), the oxygen cell, the oxygen cell fit- ting. The Measuring System  The oxygen cell interconnecting wiring, and control electronics. The Temperature System ...
  • Page 13 Figure 1-1. WDG-V Schematic. System Overview...
  • Page 14 The Oxygen Measuring Cell The sensing element itself is a closed-end tube or disk made from ceramic zir- conium oxide stabilized with an oxide of yttrium or calcium. Porous platinum coatings on the inside and outside serve as a catalyst and as electrodes. At high temperatures (generally above 1200°F/650°C), oxygen molecules coming in contact with the platinum electrodes near the sensor become ionic.
  • Page 15 where R and F are constants, T is absolute temperature, and O and O are the oxygen partial pressures on either side of the cell. For measuring oxygen in non-combustibles gases, the calibration of an analyzer is obtained from the formula: 20.9% E = A * T * Log AT = 45.78 at 650°C...
  • Page 16 Temperature Sensors There is one tightly controlled temperature zones in the WDG-V . The Oxygen Cell Temperature. Cell Temperature The Oxygen cell is controlled to 680°C. The cell temperature is measured by a dual, Type K thermocouple. A dual thermocouple is used for self check of the cell temperature.
  • Page 17 Electronics • The electronics/software completely control the sensor. - No external controllers are required. • The configuration/monitoring interface to the electronics is RS485/RS232 MODBUS RTU. • Sensor wiring and Customer wiring (i.e. power, I/O) are directly connected to the analyzer through the rear panel. - Connections are pluggable.
  • Page 18 An auto calibration can be initiated using the contact input peripheral. (Only with the RCU option). Remote Calibration Unit Interface The WDG-V has 5 solenoid valve drivers that are compatible with AMETEK’s Remote Calibration Unit (RCU). See wiring section for interface details. Communications Interface There is three communications interfaces on the WDG-V •...
  • Page 19: Common Operator Errors

    Common Operator Errors These are some common errors to avoid; they are avoided, your analyzer will operate with a minimum of maintenance and troubleshooting. Do not use pipe dope or any other contaminant that gives off combustible  vapor which may cause erroneous measurements on any joints of the sample tubing.
  • Page 20: Technical Support

    Technical Support AMETEK/Thermox is committed to providing you the best technical support in the industry. If you need service or application assistance, please call AMETEK at (412) 828-9040, or your local AMETEK/Thermox representative. Before you call the factory for technical support, run test gases and record the following val-...
  • Page 21: Chapter 2 Specifications

    SPECIFICATIONS Principle of Operation: Extractive net oxygen analyzer for use in a wet or dry gas CEM system. Zirconium oxide. Operating Range: From 0-1 to 0 to 100% Oxygen Display: 4.2” Color 1/4W VGA with Graphical User Interface (GUI). Password Protected Keypad: 18 Key membrane Accuracy:...
  • Page 22 Repeatability: ±1% of reading or 0.1% absolute, whichever is greater Maximum Inlet Temperature: 204°C (400°F) Sample Pressure: ± 2 psig max (0.14 kg/cm Sample Flow: 2 to 20 scfh (0.94 to 9.4 L/min.) Analog Output: Linear current output. The output can be 4-20 mA, 0-20 mA, 20-4 mA, 20-0 mA and is fully scalable.
  • Page 23 Calibration: Calibrate or verify calibration. Store last calibration and verification data. Se- lectable calibration gas run time and process recovery time. Timed automatic calibration with optional remote calibration unit. Environment: Ambient Temp: -25°C to 50°C Relative Humidity: 10-90%, non-condensing Max Altitude: 3000 meters Enclosure: 19”...
  • Page 24 This page intentionally left blank. 2-4 | Thermox WDG-V...
  • Page 25: Chapter 3 Installation

    INSTALLATION The operations in this chapter should be performed only by qualified service personnel experienced in electrical safety techniques. There are no operator-serviceable components inside the WDG-V system, and an operator should not attempt to open the sensor cover for any rea- son.
  • Page 26: Mechanical Installation

    Mechanical Installation This section describes how to perform the mechanical installation portion of your system installation This includes the following: • Mounting the Sensor • Sample Gas Requirements/Connections • Calibration Setup • Remote Calibration Unit Mechanical Installation (optional). Observe the following guidelines when selecting an analyzer installa- tion location: 1.
  • Page 27 Figure 3-1. WDG-V front view. Figure 3-2. WDG-V mounting dimensions. Installation...
  • Page 28: Sample Gas Requirements/Connections

    PROCESS & ANALYTICAL INSTRUMENTS DIV. IN HOUSE ONLY 150 FREEPORT ROAD REPRODUCTION PROHIBITED WITHOUT DIMENSIONS ARE IN INCHES PERMISSION BY AMETEK P & AI DIVISION PROCESS & ANALYTICAL INSTRUMENTS DIVISION PITTSBURGH, PENNSYLVANIA 15235 U.S.A. TOLERANCES: FRACTIONAL 1/32" WDG-V RM UNIT...
  • Page 29: Required Calibration Gases And Tubing

    Required Calibration Gases and Tubing Use calibration gases between 2 and 20 SCFH (0.94 to 9.4 L/min). Also be sure that you use the same flow rate for calibrations and daily operations. The span gas must be higher than the zero gas by a factor of 10 (for example, if zero gas is 1.0%, span gas must be 10% or higher.
  • Page 30: Remote Calibration Unit Mechanical Installation (Optional)

    Remote Calibration Unit Mechanical Installation (Optional) Remote calibration unit (RCU) mechanical installation includes the following: • Mounting the RCU. • Plumbing calibration gases to the RCU. • Plumbing the RCU to the sensor. See the “Wiring” section in the Installation chapter for help on how to ...
  • Page 31 Standard RCU mechanical installation Figure 3-5 shows mounting dimensions for a standard RCU. Use two #10 screws to mount the RCU. Note that only the top left and lower right holes are used. The ambient temperature range for the RCU is -20 ° to 70 °C. Mount the RCU as close to the sensor as possible.
  • Page 32 Make the following calibration gas connections to the standard RCU. Connect instrument grade air that is free of oil or dirt to the instrument air inlet connection on the left side of the RCU. If you are also using instrument air as your span gas, you can skip Step 2 below because instrument air will be used for both aspirator air and the span calibration gas.
  • Page 33: Wiring

    Wiring Remove AC mains power from the controller before performing wiring. Connections to the analyzer are made through the rear analyzer panel (see Figure 3-7). Any screw terminals not described in this section are reserved for future use. Installation...
  • Page 34 Figure 3-7a. Customer I/O Interconnecting WDG-V 3-10 | Thermox WDG-V...
  • Page 35 Figure 3-7c. Customer I/O Interconnecting WDG-V Installation 3-11...
  • Page 36: General Wiring And Conduit Requirements

    This wiring section shows you how to make the following connections: • AC mains supply wiring to sensor • Sensor to remote calibration unit • Sensor to alarm devices • Sensor to current output devices • Sensor communications In addition, this wiring section provides mandatory EMC grounding, shielding, and noise protection requirements.
  • Page 37: Sensor Ac Mains Supply Connections

    You must use twisted-pair cable in rigid metal conduit or use twisted pair cable with an overall braided shield. All cable shields or conduits connecting to the control unit must be chassis grounded. EMC Grounding Method Use properly grounded cable to connect to power entry module. Transient and RFI Interference This section describes transient and RFI interference precautions: •...
  • Page 38: Current Output Connections

    Current Output Connections Standard current outputs There are three current outputs on the WDG-V sensor. The current output con- nections are labeled as follows on the sensor board terminals: Analog Output #1 => +I1- Analog Output #2 => +I2- Analog Output #3 => +I3- The current outputs are referenced as Analog Outputs 1, 2 and 3 on the user in- terface.
  • Page 39 Figure 3-9. Output Device Max load 1000 Ω Installation 3-15...
  • Page 40: Alarm Contact Connections

    Alarm Contact Connections Follow general EMC grounding and shielding requirements for all wiring as described in the “General Wiring Requirements” section at the beginning of this Wiring section. The alarm contacts are labeled FLTA/B, DVLDA/B, ALM3A/B, ALM4A/B, ALM5A/B, on the rear panel of the analyzer. They have a rating of 10 VA, maximum 30 volts AC or DC (see Figure 3-9).
  • Page 41: Remote Calibration Unit Connections

    Remote Calibration Unit Connections If you don’t have the remote calibration unit (RCU) option, skip this section. Oxygen-only remote calibration unit connections Oxygen-Only Remote Calibration Unit (RCU) connections on the WDG-V ana- lyzer and their RCU connections, are as follows (see Figure 3-11): ZERO GAS - WDG-V Terminal Z to Pin 13 on RCU ASPIRATOR - WDG-V Terminal A to Pin 14 on RCU SPAN - WDG-V Terminal S1 to Pin 15 on RCU...
  • Page 42 Figure 3-11. RCU Connection 3-18 | Thermox WDG-V...
  • Page 43: Communications Connections

    Digital input to initiate remote calibration unit (Optional, consult factory) The digital input connections on the wiring card allow you to initiate a remote cali- bration from a location other than the analyzer. For this option to work, you must have a remote calibration unit.
  • Page 44 This page intentionally left blank. 3-20 | Thermox WDG-V...
  • Page 45: Chapter 4 User Interface

    User Interface Home screen This is the main screen after boot-up which displays concentration values, system status and alarms if present. The Home screen items include: Gas Concentrations Trend Plot (30 min) - only on O Temperatures (hidden in normal mode) Time/Date Sensor Name Last Calibrations time and date...
  • Page 46 Date sensor name time concentrations real time trend Graph (30 Minutes) Last status Bar calibration Dates figure 4-1. Home screen The bottom bar as seen in the picture above shows the Status and Alarms. This bar turns RED when in FAIL state and YELLOW when warnings are present. This Home screen is not configurable except the thermometers can be viewed in Normal Mode by pressing the UP arrow.
  • Page 47: Main Menu

    Main Menu The menus have a cover flow system similar to many smartphones. The menus wrap in either direction and can be navigated by using the right or left arrows. Press Menu to select. figure 4-2. Main Menu screen User Interface...
  • Page 48: Menu Structure

    Menu structure Show below is the complete menu tree for the user interface. figure 4-3. Menu structure 4-4 | Thermox WDG-V...
  • Page 49: Chapter 5 Configuring Your Analyzer

    CONFIGURING YOUR ANALYZER Sensor Configuration There are several options for setting up your sensor depending on your particular process and application. You can set these by selecting Sensor Configuration in the Configuration Menu. Figure 5-1. Configuration Menu Screen For the WDG-V , the only options is: Auto Calibration.
  • Page 50: Analog Outputs

    Figure 5-2. Configuration Menu Screen The WDG-V software allows for up to three configurable analog outputs. This can be setup through the configuration menu. Analog Outputs You can direct process readings to any of the three analog output ports. Configurable for: Function (O2) Cell Mv Mode (0-20, 4-20, Namur) Span and Zero...
  • Page 51 The Namur Standard is implemented as follows: 4-20 ma Analyzer Condition Signal 0 mA Analyzer unpowered, or completely failed 3.5 mA Critical Alarm - analyzer reading unusable 3.8 mA Reading Under Range (Example - user sets range to 2-10%. Current reading is 1.9%) 4 to 20 mA Normal Operation 20.5 mA...
  • Page 52 Analog Output 1 The image below shows the analog output configuration screen. Use Left/Right Arrow to navigate to the next Analog Output. Figure 5-4. Analog Out Screen Track Hold During Cal Options - Track during Cal - Hold last value during Cal 5-4 | Thermox WDG-V...
  • Page 53: Setting The Alarms

    Setting the Alarms The WDG-V has five independent, normally open alarms. The relays can be set to energize or deenergize on alarm. Alarms: Process, Service, Fault and Data Valid There are three (3) process relay alarms, (1) fault alarm and (1) data alarm. Process alarms are activated when the process reading levels, based on what the user has defined in the setup menus, have been exceeded.
  • Page 54 Alarm Configuration The image below shows the Alarm Configuration Screen Use Left/Right Arrow to navigate to the next Relay. Figure 5-6. Alarm Configuration Screen Here, set high and low alarm limits and determine whether the signal should energize or de-energize upon alarm. Figure 5-7.
  • Page 55: Communications

    Communications Configuration for TCP/IP and customer Modbus. Figure 5-8. Configuration Menu Screen TCP/IP Configuration Configurable for: IP Address Subnet Mask Gateway IP DHCP enable/disable Figure 5-9. TCP/IP Configuration Screen Configuration...
  • Page 56 Modbus Configuration Configurable for: Baud Parity Stop Bits Figure 5-10. Modbus Configuration Screen Left/Right Arrow to toggle between Modbus and TCP/IP configuration 5-8 | Thermox WDG-V...
  • Page 57: System Time

    System Time Set System Time. Set correct time for data logging and system events to log properly. Figure 5-11. Configuration Menu Screen Entering The System Time Enter the correct local date and time as shown below. Use 24 hour/military time notation for setting the hour.
  • Page 58: Passcode Configuration

    Passcode Configuration The WDG-V analyzer allows for two separate 6 digit passcodes. A “user” level passcode gives access to the menus, but will not allow for any changes to the analyzer configuration. A “super user” level passcode will give access to all menus includ- ing the analyzer configuration menu.
  • Page 59 Configuring the Passcode Enter and verify the new passcode using the keypad. The “super user” passcode should be set up first in order for the system to operate correctly. The “user” passcode will give access to the menus, but will not allow for any changes to the analyzer configuration.
  • Page 60: System Name Configuration

    System Name Configuration Figure 5-15. System Name Configuration screen. From the Main Menu screen, press the ENTER/MENU key to access the Configuration Menu. Use the left and/or right navigation arrows to locate the System Name Configuration menu and press the ENTER/MENU key. Use the keypad to enter the name(s) of your analyzer(s).
  • Page 61: Flow Meter Configuration

    Flow Meter Configuration To configure the flow meter, select Flow Config from the Configuration Menu. Figure 5-16. Flow Configuration screen. Use the drop down menu to select the untis for use on the analyzer - scfh - lpm - ccm Use the keypad to enter the low flow alarm threshold in the appropriate units.
  • Page 62: Flow Calibration Procedure

    Flow Calibration Procedure Calibration of the flow meter is only required if the sample flow is not vented to atmosphere. Performing a calibration will create an offset to compensate for any pressure effect. An accurate external flow measurement is required to perform the calibration and is not supplied with the analyzer.
  • Page 63 CALIBRATING YOUR ANALYZER Oxygen Calibration To calibrate the analyzer, select Start Calibration from the Calibration Menu. The WDG-V analyzer should be allowed to reach temperature before  starting any calibration. It is recommended that the analyzer be allowed to run for a minimum of 2 hours before starting calibration. NOTE Figure 6-1.
  • Page 64: Chapter 6 Calibration

    Calibration The Calibration Menu allows the user to configure the sensor for calibration, per- form the calibration, view the last calibration and calibration history. Start Calibration Select gases and cal gas values; start a calibration Cal Gas Values Enter cal gas values Last Cal Results Shows results for the last calibration Cal History...
  • Page 65 Cal Gas Values Menu Use this menu to set and save the calibration gas values for the sensor you have chosen. These values will be used when calibrating the sensor. This menu provides a way for the user to select the values before calibrating the analyzer. However, if the values are not set here, they can still be set during the calibration process under the Start Calibration menu.
  • Page 66 Start Calibration Menu Select Gas to be Calibrated Oxygen Recovery Duration Define a time period for the system to recover from reading calibration gas values to reading process gases. Calibration Gas Values Enter the calibration gas values to match the span and zero calibration gas cylinders you will use to calibrate or verify your system.
  • Page 67 Calibrate the Analyzer Perform a Manual Calibration From the Main Menu screen, press the ENTER/MENU key to access the Cali- bration Menu. Use the left and/or right navigation arrows to locate the Start Calibration menu and press the ENTER/MENU key. You are prompted to select the gases to be calibrated, or choose the “All”...
  • Page 68 Figure 6-3. Calibration screen gas status reading. 10. When all gases have calibrated, the screen displays the Calibration Results that include % drift as well as recovery time. If the calibration is successful, a large green check mark will be displayed. If it is not, a large red X is displayed. Figure 6-4a.
  • Page 69 Perform an Automatic Calibration The functions for automatic calibration will only be visible if you have an RCU and have configured your sensor for Automatic Calibration in the Sensor Configuration menu. Figure 6-5a. Auto Cal Config Menu Figure 6-5b. Start Calibration Menu. There are two ways to perform an automatic calibration: Define the automatic calibration time and date in the Auto-Calibration Configura- tion (Auto Cal Config) menu.
  • Page 70 Figure 6-6. Auto-Calibration Configuration screen. Cal Gas Duration Define how long you want each calibration gas to flow through the sensor before switching the gases. in Seconds The RCU automatically switches the span and zero calibration gases based on the calibration gas dura- tion times you specify.
  • Page 71 Figure 6-7. Select calibration type. Press the right arrow to continue. Select the gas(es) that you want to calibrate. Enter the time for the Cal Gas Duration in seconds. Enter the time for the Recovery Duration in minutes. Press the right arrow to continue. This is the Calibration Gas Values screen.
  • Page 72 Perform a Remote Calibration The functions for remote calibration will only be visible if you have and RCU and you have configured your sensor for Automatic Calibration  in the Sensor Configuration menu. NOTE Using the Remote Cal Unit (RCU), the user manually accepts the gas values when each gas is stable by pressing the ENTER/MENU key, but the RCU switches the gases on/off automatically.
  • Page 73: Flow Calibration

    You will be prompted to “Inject the Zero Gas.” Press the ENTER/MENU key and the RCU will apply the gas. 10. Calibration gas values and status are displayed. Press the ENTER/MENU key to accept the values. The status line at the bottom of the screen indicates which gas is being ...
  • Page 74 Flow rate will now be compensated. Calibration is valid only if the pressure remains consistent. Any sig-  nificant change in sample pressure will require re-calibration. NOTE 6-12 | Thermox WDG-V...
  • Page 75: Chapter 7 Diagnostics

    Diagnostics Diagnostics Diagnostics are used to help determine the overall health of the sensor as well as for basic troubleshooting in the unlikely event that there is a problem with the analyzer. The display screens give visibility to how the analyzer is running and where glitches might occur.
  • Page 76: Sensor Diagnostics

    sensor Diagnostics Use the left and/or right navigation arrows to locate the Sensor Diagnos- tics menu and press the EntEr/mEnu key. sensor Diagnostics screen 1 Cell Temperature degrees Celsius Box Temperature degrees Celsius Cell HDC Box HDC On Time days, hours Total on Time days, hours Use the right navigation arrow on the AMEVision enclosure to move to...
  • Page 77: Control Diagnostics

    control Diagnostics Diagnostics mode is turned on when changes are made on this menu so that the system does not accidentally trigger an auto calibration while troubleshooting. Use the left and/or right navigation arrows to locate the Control Diag- nostics menu and press the EntEr/mEnu key. When you press the EntEr/mEnu key a message appears on the Status Bar at the bottom of the screen “Diagnostic mode: OFF.”...
  • Page 78 control Diagnostics: Relays screen Figure 7-1b. Diagnostic Mode. Relays screen. This screen displays the Set Values and Actual Values for the five Outputs and allows the user to Open or Close the output relay. Use the right navigation arrow to move to the Close Contact screen. control Diagnostics: close contact screen Figure 7-1c.
  • Page 79: Inject Gas

    inject gas Use the left and/or right navigation arrows to locate the Inject Gas menu and press the EntEr/mEnu key. Use this menu to test the flow of gas to the analyzer. Use the navigation arrows to select the gas to inject and press the Enter/ mEnu key.
  • Page 80: Trend Logging

    trend Logging Plotting Data using the candlestick Format When plotting the data for trend logging, the WDG-V software uses a “candlestick” method to format the trending. This method allows more data to be plotted than there is space available on your display giving you more insight into how your analyzer is operating.
  • Page 81 The Trend Logging menu allows the user to select a gas and a period of time to plot and log how the selected gas is trending. Use the left and/or right navigation arrows to locate the Trend Logging menu and press the EntEr/mEnu key. Select the gas to log and press the EntEr/mEnu key.
  • Page 82: Cell Diagnostics

    cell Diagnostics This menu displays the information the user needs to determine the health of the cell and to determine if and when it needs to be replaced. Use the left and/or right navigation arrows to locate the Cell Diagnostics menu and press the EntEr/mEnu key.
  • Page 83: Utility

    Utility Utility Menu From the Main Menu screen, press the EntEr/mEnu key to access the Menus. Use the left and/or right navigation arrows to locate the Utility menu and press the EntEr/mEnu key. There are four utility menus: • Parameter Management •...
  • Page 84 Parameter Management This menu is used to save or restore sensor user parameters or revert the sensor parameters to factory default. Use the left and/or right navigation arrows to locate the Parameter Management menu. Press the EntEr/mEnu key to access the Sensor Parameters Manage- ment menu.
  • Page 85: Firmware Upgrade

    Check the uSB drive to make sure it is connected and try again. If you continue NOTE to fail to upgrade the firmware, contact Ametek customer service. Press the CAnCEL key to exit the menu. Diagnostics...
  • Page 86: Usb Transfer

    UsB transfer Use this menu to upload data to a USB drive for use on a computer. Use the left and/or right navigation arrows to locate the USB Transfer menu and press the EntEr/mEnu key. Press the EntEr/mEnu key to access the USB Transfer screen. The user can: •...
  • Page 87 “Failed!” You will be asked if the uSB was plugged  in. Check the uSB drive to make sure it is connected and try again. If you continue to fail to upload the trend data, contact Ametek customer NOTE service.
  • Page 88: Reset Age

    Reset age This menu allows the user to reset the age of the cell. this should only be done when the device you have changed the cell. Use the left and/or right navigation arrows to locate the Reset Age menu and press the EntEr/mEnu key.
  • Page 89: About

    about This menu option provides the user with information on the Host and Sen- sors. Use the left and/or right navigation arrows to locate the About menu and press the EntEr/mEnu key. Two menus display: • About Host • About Sensor about Host Navigate to About Host and press the EntEr/mEnu key.
  • Page 90: Alarm History

    alarm History Separate from the Trend Data, a record of all events and alarms are stored by the analyzer. This includes all passcode changes. To view the Alarm history, select Alarms from the Main Menu. Figure 7-5. Main Menu screen Then select, Alarm History.
  • Page 91: Chapter 8 Troubleshooting

    TroubleshooTing The operations in this chapter should be performed only by qualified service personnel experienced in electrical safety techniques. Les operations dans ce chapitre doivent être effectuées seulement par un technicien qualifié et expérimenté en techniques de sécurité électrique. There are no operator-serviceable components inside the WDG system. Never service the controller or sensor unless power has been removed from the control- ler and sensor, and the sensor has been allowed to cool for at least one hour.
  • Page 92: General Troubleshooting

    general Troubleshooting Your system may pass calibrations, yet still seem to be reading incorrect oxygen levels. If this is the case, you may want to check the following: leak Check Leaks can lead to inaccurate readings, especially if operating under a significant pressure or vacuum.
  • Page 93: Diagnostics Checks

    Diagnostics Checks This section shows you how to check different sensor areas for possible prob- lems. See the “Error Messages” section for help on what checks you should perform based on the error message displayed. If you don’t see an error message, yet feel your readings are inaccurate, you may also want to check the “General Troubleshooting”...
  • Page 94: Wiring Checks

    Wiring Checks Make sure all wiring connections between the control unit and the sensor are terminated to the proper locations and interconnecting Wiring Problem are seated properly. This includes ensuring that sets of wires are not reversed, and that the cable is not damaged or melted.
  • Page 95 Calibration/Aspirator setup Checks Calibration gas Check To check calibration gas  Check that the correct calibration gas values have been en- values: tered into the analyzer. Select the Cal gas Value  Check that the calibration gas cylinders are turned on and are menu option from the Cali- not empty.
  • Page 96: Ac Power Checks

    To check if a Verify that no drive signal is present and check for flow on the calibration gas RCU flow meter. If flow is indicated when no solenoids are ener- solenoid is stuck gized, a solenoid is stuck open. Shut off your calibration gases open: (one at a time) until the flow drops to zero.
  • Page 97: Process Pressure Checks

    Process Pressure Checks To check that you entered the process pressure correctly, select Process Pressure from the configuration menu. Also ensure that the calibration process pressure equals normal operat- ing process pressure. Calibration should only be performed under these conditions for highest accuracy.
  • Page 98 Alarm and Warning Messages AlArMs AlArM DesCriPTion Possible Causes Cell T/C Failure Thermocouple is not connected or is open Bad T/C, Bad connection Cold Junction Compensator The PCB temperature sensor has failed Bad Electronics Failure Cell Temperature Control A critical overtemperature condition has Shorted, open, or Bad occurred T/C or Failed Temperature...
  • Page 99 WArnings Cell Life Nearing Its End The ZrO cell is nearing its end of life. This Low cell output is based on calibration history. High Cell Mv The combustible measurement is no longer Hydrocarbon upset valid (set to full scale) due to lack of O Analog Output 1 Out Of Range The analog output value is saturated (high Improperly set range or low)
  • Page 100: Led Status Indicator

    leD sTATus inDiCATor Main status leD (located on main sensor board behind dipswitch) OFF – Power is off. The status LED is NOT directly connected to the WDG-V power supply. In the unlikely event of an error condition, it is possible for ...
  • Page 101: Chapter 9 Service And Parts

    SERVICE AND PARTS Replacement Parts List Sensor Furnace Assembly 120VAC P/N 7000-872-SE Furnace Assembly 230VAC P/N 7000-873-SE Kit, WDG-V Replacement Thermocouple for O Cell Furnace P/N 7000-839-TE Peripherals and I/O Connection Board Assembly P/N 8000-120-SE Cell O-Ring Metal P/N 42005JE Zirconia Cell P/N 7000-568-SE Main Processor Board...
  • Page 102 When ordering, provide the serial number of your analyzer to ensure proper parts are ordered: AMETEK Process & Analytical Instruments Division 150 Freeport Road Pittsburgh, PA, USA 15238 Phone: (412) 828-9040 Fax: (412) 826-0399 9-2 | Thermox WDG-V...
  • Page 103: Chapter 10 Customer Modbus Map

    Customer modbus map Customer Modbus Map 10-1...
  • Page 104 register parameter type description PARID_SYSALARM Int 32 0 - host is communicating with the sensor; Read Only 1 - host is not communicating with the sensor PARID_PASSCODE Unsigned Int 32 User pascode 000000 through 999999 Writable PARID_SUPERCODE Unsigned Int 32 Supervisor passcode 000000 through 999999 Writable PARID_OXYGEN Float...

Table of Contents