Service Alarm; Watchdog Relay - Ametek Thermox Series 2000 WDG-IV User Manual

Table of Contents

Advertisement

4-26 | Thermox Series 2000 WDG-IV / IVC / IVCM

Service Alarm

The Service Alarm operates as normally closed (service alarm condi-
tion causes open contacts).
NOTE
The Service Alarm provides an alarm to alert you of possible system prob-
lems that require attention. Events that can trigger the service alarm to
activate are:
Error condition (detected by software)
System operational variance (such as an over-temperature or under-
temperature condition)
Calibration failure
Verification failure
Control unit power failure

Watchdog Relay

During normal system operation, the Watchdog Relay operates as
normally closed (system failure causes open contacts).
NOTE
The Watchdog Alarm relay de-energizes when the system watchdog timer
times out. The watchdog timer will time out and cause the control unit to
reset itself to prevent the control unit from becoming "lost" due to such
things as noise spikes or power surges. You can use this relay to activate a
remote indicator to notify you if the watchdog timer times out.
In addition, the control unit reads the watchdog after reset, and beeps for
a brief time if the reset was caused by a watchdog condition. Pressing the
reset keys simultaneously (hidden keys to the left and right of the Setup
key) for a long enough time will cause a watchdog. This will indicate as a
beep after the system restarts after releasing the reset keys.

Advertisement

Table of Contents
loading

Table of Contents